Electrical Connections
Electrical Shock Hazard – All input and output wiring should be installed in such a manner that
access to bare copper is prohibited during normal operation .
An access hole sized for ½" NPT (15 mm) electrical conduit fitting is provided on the
bottom of the enclosure. A sealing type conduit fitting is required to maintain the
enclosure's rating.
Power Connection
The Sample Sentry II contains two circuit boards stacked in the enclosure. The power
connection is located on the output board, the bottom board in the circuit board stack.
The Sample Sentry II contains a universal power supply. The power supply is capable
of accepting an input voltage range of 100–240 VAC and frequency of 50 or 60 Hz. The
Sample Sentry II automatically operates with power supplied in this range. A power supply
lead of 16 AWG (1.3 mm²) copper wire is recommended.
The Sample Sentry II should be connected to an electrical outlet or outlet strip with surge
suppression and filtering. A suitable external over-current protection device such as a fuse
or circuit breaker (15 A) is recommended for both the hot (L) and neutral (N) leads. Using
this method, the plug serves as the disconnect device. The cord and plug shall comply
with the following standards:
Table 2. Power Cord Requirements
Location
Cord
United States
ANSI/UL 817
Canada
CSA C22.2 No. 21 or C22.2 No. 49
European
IEC 60227 or 60245
Community
If desired, the Sample Sentry II may also be hard wired to instrument-quality power
using appropriate certified conduit, fittings, and wiring. A suitable external over-current
protection device, such as a fuse or circuit breaker (15 A), and disconnect device is
recommended. The over-current protection and disconnect devices shall be installed on
both the hot (L) and neutral (N) leads. The disconnect device shall be located near the
equipment and marked with appropriate ON(1) OFF(O) markings as specified by local
codes. For fixed wiring methods, the ground conductor shall have green with a yellow
stripe as the insulation. Installation should be performed by qualified personnel in
accordance with local codes and procedures.
Sentry Sample Sentry II
Plug
ANSI/UL 498
CSA C22.2 No. 21 and C22.2 No. 42
Appropriate IEC standard