Measuring Ac And Dc Current - SEFRAM 7352B User Manual

Trms ac+dc bluetooth digital multimeters
Table of Contents

Advertisement

Available languages

Available languages

Measuring AC and DC Current

 WARNING
Always use proper terminals, switch position, and range for
measurements.
To avoid possible electric shock or personal injury, never
attempt an in-circuit current measurement where the
open circuit potential to earth is greater than 1000V.
Only replace the blown fuse with the proper rating as
specified in this manual.
 CAUTION
Replace the fuse as soon as the indicator (FUSE) appears.
To measure current, you must break the circuit under test, then
place the meter in series with the circuit.
The ranges of measuring current are 40mA, 400mA, 4A and
10A. AC current is displayed as an rms value. Insert the black
lead into the COM terminal. For currents less than 400 mA,
insert the red lead into the mA terminal. For currents above 400
mA, insert the red lead into the A terminal.
You can press the function (yellow) button to select current
measurement function.
8

Advertisement

Table of Contents
loading

Table of Contents