Visitor T30 C-DR User's Manual CREDITS Compliance This medical device is in compliance with the Medical Device Directive 93/42/EC and its revised versions. The medical device, hereafter called equipment, has been classified in class IIb according to annex IX rule 10 of the directive mentioned above.
User's Manual Visitor T30 C-DR Information about User's Manual The purpose of this User's Manual is to provide a valid help in order to ensure a safe and efficient use of the described equipment to the users. Before starting up the equipment, it is necessary to read the User's Manual, note down and strictly respect all the notices indicating Warning and Precaution messages.
Visitor T30 C-DR User's Manual Compatibility The equipment described in this User's Manual mustn't be used together with other products or components, except in case they are explicitly indicated as compatible by the Manufacturer. A list of these products and components is available by the Manufacturer.
User's Manual Visitor T30 C-DR SAFETY Warnings and precautions Maintenance and defects Do not use the equipment for any application before the user correctly performs all regular checks and updates the periodical equipment maintenance. If it is sure ( or probable) that any part of the equipment is defective or wrong adjusted, don't use it before performing all reparations.
Page 9
Visitor T30 C-DR User's Manual The Manufacturer is responsible for safety features of its own products, only provided that maintenance, repairs and modifications are performed exclusively by the Manufacturer's personnel or by personnel expressly authorized by the Manufacturer. As for all technical equipments, even this medical device must be used properly and subject to regular maintenance and care, as described in "...
User's Manual Visitor T30 C-DR Electrical Safety This equipment is in compliance with safety class I, Type B, in accordance with IEC 60601-1 norm. Do not use the equipment near or leaned against other equipments. Do not remove protections or cables from this equipment, unless it is expressly required in this User's Manual, because inside it there are dangerous electrical voltages.
Visitor T30 C-DR User's Manual Protection against explosions This device mustn't be used in presence of explosive gas or fumes, such as some kind of gaseous anesthetics. Do not use disinfectant spray flammable or potentially explosive. The use of this equipment in an unsuitable environment can cause fires or explosions.
User's Manual Visitor T30 C-DR Electromagnetic compatibility (EMC) This equipment complies with international and national laws and regulations relating to electromagnetic compatibility (EMC) in force for this type of product, if it is used for the intended purposes. Such laws and regulations define the electromagnetic emissions level coming from the product and the requested immunity against electromagnetic interferences from external sources.
Visitor T30 C-DR User's Manual 2.7.1 Warning and Safety Precautions for Electromagnetic Compatibility Increased emission or reduced interference immunity. Use of unsuitable accessory or lines ► Exclusive use of the listed accessory or line with the exception of internal original spare part components.
Page 14
User's Manual Visitor T30 C-DR Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
Visitor T30 C-DR User's Manual Protection against ionizing radiations This equipment generates ionizing radiations ( hereinafter called radiations). Before proceeding with x-ray exposure, be sure that all safety measures in protection against radiations have been taken. While using the equipment, the examination room personnel have to respect all necessary protection rules.
User's Manual Visitor T30 C-DR 2.10 FPD,warnings and precautions Risk of damages to people Anomalous use of the FPD component. ► Don't use the component for uses other than those foreseen. ► Don't use the component as main shield against x-rays.
Visitor T30 C-DR User's Manual 2.11 Labeling (A) Equipotential node (B) Potter connections (C) Connection of hand switch for exposures (D) Magnetothermic switch (A) Label for the monobloc rotation lock. By rotating the knob clockwise, the rotation stops. (A) Label for the parking brake lock.
Page 18
User's Manual Visitor T30 C-DR (A) Label for the movement position of the equipment. (A) Label for the movement position of the equipment (B) Label for the exposure hand switch (A) Laser openings (B) Warning labels Avoid the exposure Radiations are emitted...
Page 19
Visitor T30 C-DR User's Manual Laser radiation symbol (B) Warning label: Laser Radiation Do not stare into beam Class 2 LASER product IEC 60825-1:2007 Po ≤ 1mW; λ = 645 ±10nm [File: 218899-01-01-B.docx] Page 19/74...
User's Manual Visitor T30 C-DR MAINTENANCE, CLEANLINESS AND DISPOSAL Periodic maintenance For this equipment it is necessary : a correct operation, a periodic maintenance and checks that the user must regularly perform and that are essential for the equipment to work safely, efficiently and reliably.
Visitor T30 C-DR User's Manual Cleaning and disinfection Only personnel trained in the management of cleaning and disinfection of medical devices is authorized to conduct such activities. Perform regularly cleaning and disinfection operations of the equipment. Below are the instructions.
User's Manual Visitor T30 C-DR Disposal The manufacturer wants to make a contribution to environment defense and wants to guarantee a constantly safe and efficient use of this equipment by using a proper support, maintenance and training program. If the equipment is used correctly and always subjected to proper maintenance, it doesn't represent an environmental risk.
Visitor T30 C-DR User's Manual COMPONENTS IDENTIFICATION Usability The equipment is used in hospital environment to perform x-ray examinations, in particular in cases when the transport of the patient in a ward with fixed equipment is uncomfortable or not possible.
User's Manual Visitor T30 C-DR Overview A - Monobloc B - Goniometer C - Collimator D - Monobloc handle E - Cable reel F - Power supply cable G - Circuit breaker/ Equipotential node H - Exposure control hand switch...
Visitor T30 C-DR User's Manual Safety devices Circuit breaker The equipment is protected by a circuit breaker (A) against excessive mains fluctuations. In case of intervention of the circuit breaker, to restore the working of the equipment it is enough to put back the control of the circuit breaker in "I"...
User's Manual Visitor T30 C-DR X-ray generator control panel All keys are membrane type Symbol Function Description Equipment OFF Equipment ON. The green led indicates that the equipment is connected to the mains and powered. POTTER POTTER selection. The green led ON indicates the performed selection.
Visitor T30 C-DR User's Manual Symbol Function Description Function keys: F1, F2, F3, F4. They refer to the display line number (display with 4 lines). Inside the APR program, each pushbutton refers to the relative display line, by choosing the described function.
User's Manual Visitor T30 C-DR 4.10 "Digital Imaging System" operating screen 4.10.1 DeLuxe The operating screen appears at the opening of an exam. It is divided into two sections: Frames Area and Commands Area. The Frames Area consists in acquired frames and it contains all data concerning the exam.
Visitor T30 C-DR User's Manual MESSAGES The equipment foresees three types of messages on display: S = Status of equipment No influence on the working W = Warning Press the key RESET on the keyboard to delete the warning message and go on to work.
Page 30
User's Manual Visitor T30 C-DR Error messages Typ. Text Meaning Intervention FAULTY POWER Error on the Charger or Chopper. Switch OFF, wait some minutes, Energy not available switch ON, if the error appears gain call Service ERROR V3 Absent V3 power supply...
Visitor T30 C-DR User's Manual Digital Imaging System signal and messages 5.1.1 DeLuxe In the area of messages indication (A) the status and alarm messages concerning the equipment working are displayed: FPD panel status Wi-Fi signal status/quality X-ray generator status Battery charge status of FPD panel In case of alarm, press on RESET pushbutton.
User's Manual Visitor T30 C-DR TRANSPORT Tilting hazard. Use and transport on inclined floors. ► Don’t use the equipment on floors with inclination higher than 5°. ► Don't move the equipment on floors with inclination more than 10°. Danger of damages and injuries during the equipment movement.
Visitor T30 C-DR User's Manual OPERATION Start up 1. Unroll the mains cable from the cable reel and extend it completely. 2. Insert the plug in a standard wall outlet by keeping the cable extended. The presence of the mains voltage is indicated by the switching-ON of the green led near the ON push button (A).
Page 34
User's Manual Visitor T30 C-DR 7. Charge phase of the capacitors group. The display indicates "BUSY". The phase of equipment start-up changes from a few seconds to two minutes, according to the residual charge of the capacitors. 8. The writing "BUSY" is replaced by the writing "READY".
Visitor T30 C-DR User's Manual Digital Imaging System switching ON 1. Press the Pc ON key (A) 2. Wait for the start of the operating system and application The switching-ON time of the operating system and application is about three minutes.
User's Manual Visitor T30 C-DR Positioning 1. Place the monobloc–collimator group (A) on the interested patient part. The monobloc-collimator group can rotate of +102° and -49° in respect of its axis (B) and around the arm axis of ±180° (C), 2.
Visitor T30 C-DR User's Manual Exposures The values indicated in the following images are only indicative. Exposures in DR mode 7.5.1 DeLuxe From the window of exams management, it is possible: create a local archive for a new patient (A)
User's Manual Visitor T30 C-DR Exposure in APR "digital" mode The exams preset in "digital" APR are the same preset in standard APR radiography for number and x-ray data. It is possible to create new exams and projections and change name and data of the exams already present with the intervention of qualified and trained service.
Visitor T30 C-DR User's Manual Exposures in CR mode 7.7.1 DeLuxe 1. Press the key (A) to pass from the DR mode to the CR mode and vice versa. In CR mode it is possible to perform exposures on film in "Free exposure"...
Page 40
User's Manual Visitor T30 C-DR Every group is composed by six APR programs. The list of the six programs covers three pages. The following or previous page is underlined with the symbol "▼" or "▲". 3. Press MENU key to scroll inside the pages of the chosen group.
Visitor T30 C-DR User's Manual 7.7.4 APR data table APR programs suggested and preloaded by the manufacturer on the equipment, only represent recommendations to be applied to the patient in order to optimize the operation and result of the examination.
Visitor T30 C-DR User's Manual Operation mode The equipment manages two exposure operation modes that can be selected by the operator at any minute: LOCAL: the exposure is performed with the wire control with extensible cable supplied with the equipment.
Page 44
User's Manual Visitor T30 C-DR REMOTE mode (optional) Ionizing radiations. The operations described below require the emission of ionizing radiations. ► Take the proper measures in order to avoid exposing any part of the body to direct or indirect radiations.
Visitor T30 C-DR User's Manual Indication of discharged batteries The indicator of discharged batteries (red LED) flashes with 1 flash per second when the batteries have a limited autonomy (recommended replacement) and with 2 close flashes every second when the batteries are discharged (mandatory replacement).
User's Manual Visitor T30 C-DR Optional: radiography with examination table or Potter Bucky grid The examination table or the Potter Bucky must be in compliance with IEC 60601-1 standard and must be connected to the equipment according to 93/42/CE medical devices directive and following revisions.
Visitor T30 C-DR User's Manual 7.10 Optional : DAP meter The installation and maintenance of the DAP meter can be performed only by authorized service personnel. To correctly determine the dose value to which the patient has been exposed, it is necessary to reset the dose value before starting the exam.
User's Manual Visitor T30 C-DR 7.11 Use end Potential risks Damages to the connectors and equipment. ► Grip the plug to extract the connector ► Don't extract the plugs from the outlets by pulling the cable. ► Handle the connectors with care.
Visitor T30 C-DR User's Manual TECHNICAL SPECIFICATIONS Electrical data Description Data Power supply 115 ÷ 230Vac ±10%, standard monophase with ground conductor. Automatic setting of the equipment according to mains voltage Frequency 50/60 Hz ± 5 Hz Absorbed current 10 A...
User's Manual Visitor T30 C-DR Mechanical data Description Data Width (in transport position) 618 mm Length (in transport position) 1366 mm Height (in transport position) 1483 mm Height of the transport handle 949 mm Description Data Focus-floor distance (Z-axis) 417 mm ÷ 2092 mm Max.
Page 51
Visitor T30 C-DR User's Manual All dimensions are in mm. Linear tolerances ±5 mm, angular ±1°. [File: 218899-01-01-B.docx] Page 51/74...
User's Manual Visitor T30 C-DR Operation data Description Data User interface Keyboard with alphanumeric LCD display 4 lines x 20 characters for all operative parameters and messages of possible anomalous conditions Service program for the management of the errors and the faults...
User's Manual Visitor T30 C-DR X-ray group 8.7.1 Monobloc Rotating anode monobloc model MHF2030 HV generator Thermo-tinned structure Max. power (100 kV - 320 mA) 32 kW Max. voltage to the tube 125 kV Max. current to the tube 450 mA Ripple to the max.
User's Manual Visitor T30 C-DR Collimator Description Data Model R108 F Collimation Manual with internal light source, multilayer, squared field. Assembly plan from focus 80 mm (3.14") Coverage of the field at 100cm FFD (SID) min 0 x 0 cm, max 43 x 43 cm Lighting source Clusters of high-brightness LED power.
Visitor T30 C-DR User's Manual 8.12 Digital Imaging System 8.12.1 Software features Description Data Model DeLuxe Dynamic storage RAM, standard, 2 GB Acquisition 16 bit Post processing LUTdedicated for each exam Spatial filters Harmonization V/H Image inversion Image rotation 90°...
Visitor T30 C-DR User's Manual 8.13 Optional: DAP Meter Description Data Model Diamentor CI-P Type Device for the area-dose product measurement in x- ray diagnostics according to IEC 60580 standard. Principle of measurement Radiation measure with ionization chamber Measured quantity Area-dose product 0,01 μGym2...
User's Manual Visitor T30 C-DR 8.14 Optional: Exposures remote control Description Data Type Infrared x-ray control device (IR) Transmitter Description Data Construction ABS case with matt surfaces Durable fireproof, high resistance to bad use or falls, resistance to water, oils, organic acids and alcohol Technology Infrared light (940 nm) with carrier 38kHz.
Visitor T30 C-DR User's Manual 8.15 Labels A - Equipment serial number B - Monobloc serial number C - Collimator serial number A - Distributor's logo B - Name and address of the Distributor and the Manufacturer C - Equipment model...
User's Manual Visitor T30 C-DR ANNEXES Replacement of the wireless FPD panel battery 9.1.1 Detector model DRX-1 1. Place a tool (ball pen type) in the closing mechanism of the panel. The battery is released for an easy replacement 2. Take a fully charged battery from the Battery Charger (the full recharge is indicated by the switching ON of all green leds).
Visitor T30 C-DR User's Manual 9.1.2 Detector model 3543 EZ 1. Switch the panel OFF. 2. Rotate the closing device (A) counterclockwise and remove the battery (B). 3. Take a fully charged battery from the Battery Charger (the complete recharge is indicated by the switching ON of a green led), place the battery and rotate clockwise the closing device (A).
User's Manual Visitor T30 C-DR 9.1.3 FDX3543RPW model 1. Connect the power supply to the connector placed on the back of the Battery Charger 2. Insert the discharged battery in the Battery Charger. Ensure that the battery insertion makes a click.
Visitor T30 C-DR User's Manual Protectio box for FPD panel (optional) The wireless FPD can be provided with a protection box, for a better protection against shocks. The panel and its box can be stored together or separated, in the proper unit housing.
User's Manual Visitor T30 C-DR CONFIGURATIONS AND ACCESSORIES Description X-ray emission control with extensible cable Standard Interface for examination table or Potter Bucky Standard Remote control for x-ray emission Optional DAP meter with ionizing chamber Optional Protective casing with grid for detector...