7 Electrical connection

7.1

Cabling

Use screened two-core
instrumentation cable for the current
output of the LUC-T10, LUC-T20. For
optimal protection against
electromagnetic interference, the
screen should be grounded in the
controlroom or the nearest earthing
point. A good connection to ground is
essential to good screening.
Under certain circumstances, the
digital communication signal may be
affected if unscreened cable is used.
12
7.2

Connection diagrams

" LUC-T10, LUC-T20:
2-wire "loop-powered"
# LUC-T20:
4-wire, including mains power supply
$ LUC-T30:
4-wire, including mains power supply
#
LUC-T20
4-Wire, including mains power supply
4 mA ... 20 mA
+
L+
L-
4 mA
20 mA
I+
N
I-
L1
3
1
2
4
5
+
+
V
H
18 V ... 36 V DC
L + /L
180 V ... 250 V AC
1
90 V ... 127 V AC
L–/N
"
LUC-T10, LUC-T20
+
$
LUC-T30
4-Wire, including mains power supply
18 V ... 36 V DC, 180 V ... 250 V AC
90 V ... 127 V AC
L + /L
!
L–/N
L+
L1
1
+
V
H
+
4
mA ...
20 mA
+
!
4
5
+
V
H
+
V
H
+
4 mA ... 20 mA
d2
d4
1
+
L-
4 mA
20 mA
I+
I-
N
!
2 3
5
4
+
V
H
+
V
H