Components
Moisture Sensor (optional)
The moisture sensor protects downstream analyzers against humidity. This monitoring device reliably
signals a condensate breakthrough in case of failure of cooling devices or failure of condensate drains,
thus avoiding costly downtime and high repair costs. In case of an alarm, sample gas pumps can be
switched off and shut-off valves can be activated. The moisture sensor operates according to the principle
of electric conductivity. Its design ensures that any condensate in the sample gas flow gets directly to the
sensor surface by gravitational force. Even smallest amounts of liquid will trigger a fast and reliable
alarm. Control/alert served by an appropriate electronic control.
FEATURES MS
Protection against condensate break-through
Reliable detection of smallest amounts of liquid
Line break monitoring
High chemical resistance
Optional electronic control
Electronic Control Unit (optional)
The electronic control unit serves moisture sensors and light barriers of flowmeters and analyses their
signals. The electronic control unit is equipped with a potential-free directional contact, which triggers an
alarm notification or which switches e.g. a sample gas pump or a shut-off valve. The switch-point of the
electronic control unit can be adjusted by a potentiometer. Standard setup: Function type FS 0,5s release
delay, responsivity 75 KΩ.
FEATURES EC72.01
Control/alarm for moisture sensor
Control alarm for flowmeter light barrier
Potential-free directional contact
Adjustable switch-point
LED for operation/alarm notification
AGT-PSG GmbH & Co.KG