Modify settings of I-PLUG multi-use :
iPlug MANAGER allows you to modify all the pre-programmed setting and choose your
own.
Just modify all the parameters you need, then press the Save settings button (Pic.1). You can
change hereafter parameters (Pic.1)
-
trip duration
-
sample frequency size
-
Start delay
-
Temperature alarms
Restarting I-PLUG pdf multi-use :
If the device is in a STOP state, iPlug MANAGER allows you to reset I-PLUG multi-use for
as many time as you want. By default, it will read the last configuration used.
You can modify the parameters again (if needed), then just check the Reset device flag
(Pic.1), then press the Save settings button (Pic.1) : green LED will flash 1 time to to indicate
that new settings have been recorded.
The device will be in a READY state again.
Battery is NOT changeable by user, please bring the device back to your distributor once the
battery runs out of power for recycling.
IMPORTANT: When you reset I-PLUG multi-use, all the previous recording will be
erased. If you need to store them, be sure, before resetting, to upload them in the web-
application (press Read Temperatures).
Fev, 2017
I-Plug multi-use pdf
I-PLUG Multiuse pdf
Pic.4
Page 9 of 10
Need help?
Do you have a question about the I-PLUG pdf Multi-use and is the answer not in the manual?
Questions and answers