Technical Specification

Temperature range
Temperature control stability at 450ºC
Temperature uniformity at 450ºC
Internal dimensions
Loading capacity
Typical heat up time;
Compressed air supply pressure
Air consumption:-
Nominal Heater Power
Heater type
Control system
Electrical safety protection
Fluidised medium
External dimensions
Nett weight (less fluidised medium)
Gross weight (less fluidised medium)
Electrical supply
length
width
depth
working depth
maximum media depth
-1/3 of vol.
-1/3 of C.S.Area
20ºC to 450ºC
20ºC to 600ºC
maximum
At 450ºC
height
width
depth
20
50ºC to 600ºC
+5ºC
10ºC
500mm
220mm
515mm
330mm
395mm from porous plate to top
of media when fully loaded and
fluidised.
12 litre
2
36,666mm
180 minutes
350 minutes
204-1030kPa (30-150psi)
220 l/min
150 l/min
9kW
4 semi embedded heating
elements
3 term PID temperature
controller with thermocouple
sensor and solid state relay
Control circuit fuse
Power fuses
60kg Aluminium Oxide, Part No
F0885
930mm
1155mm
660mm
272kg
322kg
See rating plate for details