Calibration Menu; Zero Calibration; Alarm History - Helio MAP 50 Operating Manual

Table of Contents

Advertisement

C
M
ALIBRATION
Menu Selection
-General Settings
-Read Detectors
.
-Alarm Relays
>>-Calibration Menu

-Alarm History

Up & down for Menu, -> Selects
C
D
HOOSE
ETECTOR

Calibration Menu

>>Choose Detector:
Gas:
Oxygen
Set Zero:
Set Span:
Transmit:
Reading:
M=Exit > = Scroll Up/Down Adjusts
Z
C
ERO
ALIBRATION
Calibration Menu
Choose Detector:
Gas:
Oxygen
>>Set Zero:
Set Span:
Transmit:
Reading:
M=Exit > = Scroll Up/Down Adjusts
HelioGD MAP50 VER. 1.0
For the latest documentation please visit
ENU
1
0.0
1.00
Complete
20.9%Vol
1
0.0
1.00
Complete
20.9%Vol
www.HelioGd.com
 Press
or
Outputs menu
 Press
to enter menu
 Press M to return to normal operation
 Press
to select option
 Press
or
 Press
to go back
Choose the Detector to calibrate.
 Press
to select option
 Press
or
 Press
to go back
Adjust the Zero to the desired value
After transmitting the value, the
detector will save it and the reading
will be updated accordingly.
or email Sales@HelioGD.com
to scroll to Analog
to choose the relay
to choose the relay
17