Summary of Contents for ScienTECH ASTRAL AC2500

  • Page 1 (217) 352-9330 | Click HERE Find the Scientech Astral at our website:...
  • Page 2 SERIES CALORIMETERS Setup and Operating Procedures PN9458L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 3: Table Of Contents

    Thank you for choosing a Scientech Astral calorimeter. Scientech, an ISO 9001 registered company, and our employees are pleased to provide you with a product designed for years of reliable service. Please read this manual completely before using your indicator. This information will enable you to fully utilize the equipment and should be located nearby for reference.
  • Page 4: Astral™ Calorimeter Specifications

    Note 14 Precision < 1 % ± 3 % Accuracy Response Time 3 sec when connected to a Scientech Indicator in Watts Mode Dimensions DxL - inches 3.75 x 2.2 3.75 x 2.2 3.75 x 3.82 3.75 x 2.2 3.75 x 3.82 3.75 x 2.2...
  • Page 5: Absorption Of Hd Absorbing Material

    Model 301-019, is available for this purpose. A 3 meter mini DIN interconnect cable is also included. If you are using a Scientech indicator, connect the calorimeter to the indicator with the interconnect cable. Follow the detailed set up instructions that are in the indicator’s instruction manual.
  • Page 6: Correcting Astral Series Hd Calorimeters Operating Parameters For Use At Different Wavelengths

    IFFERENT AVELENGTHS Scientech Astral calorimeters in general have a flat response to all wavelengths within their specified spectral response. HD and HDX calorimeters are an exception to that rule and are calibrated at a specific wavelength by adjusting the calorimeter’s gain circuitry for that wavelength. The calibration wavelength is recorded in the Operating Parameters section at the front of the manual and on the detector’s serial tag.
  • Page 7: Operation Of Astral Calorimeters With A Digital Volt Meter

    Operation of Astral Calorimeters with a Digital Volt Meter: The calorimeters may be used with any digital volt meter (DVM) capable of reading 5 volts full scale. Connect the output of the calorimeter to the DVM. Select the DC volts mode. Direct the laser beam on to the absorbing surface of the calorimeter.
  • Page 8: Numerical Integration

    Numerical Integration: Finding the area under the curve in figure 7 is the equivalent procedure for determining pulse energy. Choose an appropriate time interval, dt, and perform the summation: E = ∑ W xdt = (dt/S)∑ V The error caused by this procedure is: dE = (dt/S) Σ...
  • Page 9: Calibration Of Astral Calorimeters Using Electric Substitution Heating

    Compare this resistance to R c in the calibration data in the front of the manual. The two should agree within 2%. If not contact Scientech. D. Remove the DVM. Connect a power supply to the SUB and HTR test points and connect the DVM to monitor the power supply.
  • Page 10: Hd Absorption Vs. Wavelength

    HD A BSORPTION VS AVELENGTH Wavelength Absorption Wavelength Absorption Wavelength Absorption µm µm µm 0.200 93.00 0.480 91.54 0.735 90.77 0.210 91.00 0.485 91.63 0.740 91.22 0.220 88.00 0.490 91.74 0.745 90.79 0.230 85.00 0.495 91.67 0.750 90.80 0.240 82.00 0.500 91.63 0.755...
  • Page 11 Wavelength Absorption Wavelength Absorption Wavelength Absorption µm µm µm 0.990 90.51 1.255 88.20 1.520 88.50 0.995 90.52 1.260 88.20 1.525 88.52 1.000 90.49 1.265 88.21 1.530 88.51 1.005 90.48 1.270 88.26 1.535 88.49 1.010 90.49 1.275 88.15 1.540 88.48 1.015 90.50 1.280 88.12...
  • Page 12 Wavelength Absorption Wavelength Absorption Wavelength Absorption µm µm µm 1.785 87.73 2.050 86.86 2.315 86.25 1.790 87.64 2.055 86.83 2.320 86.07 1.795 87.62 2.060 86.72 2.325 86.04 1.800 87.66 2.065 86.71 2.330 86.01 1.805 87.64 2.070 86.71 2.335 85.97 1.810 87.59 2.075 86.60...
  • Page 13: Factory Recalibration

    All Scientech Laser Power and Energy Measurement Systems are warranted against defects in materials and workmanship for two (2) years from date of delivery. During the warranty period, Scientech will repair, or at its option replace at no charge, components that prove to be defective. The equipment must be returned, shipping prepaid, to Scientech's product service facility.