  • Page 1 100mm and 200mm Calorimeters Setup and Operating Procedures PN11434C...
  • Page 2: Table Of Contents

    Thank you for choosing a Scientech large aperture calorimeter. Scientech, an ISO 9001 registered company, and our employees are pleased to provide you with a product designed for years of reliable service. Please read this manual completely before using your calorimeter. This information will enable you to fully utilize the equipment and should be located nearby for reference.
  • Page 3 Note 10 Note 9 Precision < 1% Accuracy Response Time 5 sec when connected to a Scientech Indicator in Watts Mode Dimensions DxL - inches/cm 6.00 x 8.00/15.24 x 20.32 9.00 x 10.00/22.86 x 25.40 Weight - pounds/kgs 6/2.72 16.27/7.26...
  • Page 4: Operation Of Large Aperture Calorimeters With A Digital Volt Meter

    Mount the post to your optical bench or working surface. If you are using a Scientech indicator, connect the calorimeter to the interface module and the interface module to the indicator. Follow the detailed set up instructions that are in the indicator’s instruction manual. If you are not using a Scientech indicator operating requirements are contained in this manual.
  • Page 5: Operation Of Large Aperture Calorimeters With An Analog Chart Recorder

    Operation of Large Aperture Calorimeters with an Analog Chart Recorder: Calorimeter Response: The response of a calorimeter to a single pulse input as displayed by a chart recorder appears below. The output voltage from a chart recorder can be converted to wattage at any time by: W = V/S, W V = Chart recorder voltage level in mV S = Calorimeter sensitivity in mV/W...
  • Page 6: Initial Voltage Interpolation

    Compare this resistance to R c in the calibration data in the front of the manual. The two should agree within 2%. If not, contact Scientech. C. Calculate the voltage equivalent to laser power using the following formula:...
  • Page 7: Limited Warranty

    All Scientech Laser Power and Energy Measurement Systems are warranted against defects in materials and workmanship for two (2) years from date of delivery. During the warranty period, Scientech will repair, or at its option replace at no charge, components that prove to be defective. The equipment must be returned, shipping prepaid, to Scientech's product service facility.
  • Page 8: Disposal Of Electrical And Electronic Equipment

    LECTRONIC QUIPMENT Scientech, Inc. recommends the following for disposal of electrical and electronic equipment: 1. The best option is to reuse the equipment in its entirety. 2. Where the equipment can not be reused in its entirety, priority should be given to reuse of its subassemblies and components.