SCP SCIENCE DigiPREP Service Manual page 52

Keypad controller
Table of Contents

Advertisement

The Dual-Zone Tube Rack may be placed on the system. The fi rst
set of cut outs is to support the Dual-Zone Tube Rack when it is in
the cooling position. The second set supports the Exhaust Manifold
module. The manifold may be placed on the top two upper cut outs;
this is the resting place of the manifold when not in use.
The drip tray is shown in place. It may be stored on top of the Tefl on
manifold when not in use.
Connect the black tubing to the Tefl on drain tubes of each of the
manifold drains from the top part of the 'Y' connectors.
The long black tubing is connected to the bottom of the last 'Y'
connector. One 'Y' is supplied for the HT 250-20 and 3 'Y' are
supplied for the HT 100-40 position. (As shown in picture). The HT
100-10 and HT 250-10 position Tefl on manifolds are connected
directly to the black exhaust tubing . The Tefl on manifold may
have 1, 2, or 4 drain tube connections. The end of the long black
drain tube connects to the water jet (supplied with system) or the
Universal Fume Scrubber (UFS option)
KeyPad
52
Version 1.0
DigiPREP HT system may now be placed in the fume hood.
The picture shown on the left shows a typical digestion set up, the insert
rack with digestion tubes in place, lowered onto the graphite heating block,
the Tefl on manifold has been lowered onto the tubes, the exhaust tube is
connected to a vacuum system for the removal of the unwanted fumes.
SCP SCIENCE