X. Appendix A: Computer Setup
The BC1200 has the capability to switch between 2 different calibration
factors. To change between the calibrations remove the computer from
the flow probe head by twisting 45 degrees counter clockwise and lifting.
The indented gray button in the upper left corner on the back is to switch
between CAL I and CAL II. In the upper left corner of the display, I is
displayed for CAL I and II is displayed for CAL II.
Note:
I = ft/sec, calibration # = 0053
II = m/sec, calibration # = 0016
The indented gray button on the upper right is used to enter the calibration
mode. Press and hold it for 5 seconds to enter calibration mode.
TO RESET THE CALIBRATION:
(Calibration #'s are factory set. Resetting is only required after
changing the battery.)
• Press bottom button until CLOCK or TOTALODO is not displayed on
screen.
• Press the left indented gray button to select CAL I.
• Turn computer over and press and hold the right indented gray button
for 5 seconds and "set language" flashes on display.
• Press top button to select language.
• Press bottom button to accept.
• Press top button until "SET M" is displayed.
• Press bottom button to accept. The calibration factor is now displayed.
• Pressing the top button will change the value of the flashing digit.
• Pressing the bottom button will accept this value and move to the next
digit.
• Set the calibration factors as follows:
o Feet/second:
o Meters/second: 0016 (CAL II)
• Press indented right button on back for one second to store.
Repeat above procedure for Cal II. (Only the cal number will be required)
Global Water
800-876-1172 • globalw.com
0053 (CAL I)
13