Classification
Electric Shock Protection
Applied Parts
Ingress Protection
Mode of Operation
Specifications
Mechanical
Water capacity
Liters (US Gallons)
Water flow rate per
turbine
Electrical
Input voltage
Line Frequency
Rated Current
Electromagnetic
Compatibility
Description
Radiated Emissions
Conducted Emissions, AC Mains
Harmonic Emissions
Flicker
Electro‐Static Discharge (ESD)
Radiated RF EM fields
Electrical Fast Transient (EFT)
Surge
RF Conducted Immunity
Magnetic Field
Short Interruption
Environmental Conditions
Operational
Transportation and
Storage
Relative humidity
Atmospheric Pressure
Thermometer
Range
Accuracy
WHITEHALL MANUFACTURING P.O. BOX 3527 City of Industry, CA 91744-0527 U.S.A.
Phone (800) 782-7706 (626) 968-6681 Fax (626) 855-4862 https://www.whitehallmfg.com
Class I
None
IP24
Continuous
F-260
1173 (310)
1382 (365)
0-56.8 L/min (0-15 US gallons/min)
115/230 V~
230 V~ models are for export only
50/60 Hz
6.9/3.5 A
All Whirlpool models comply with the latest international
electromagnetic compatibility standard (IEC 60601-1-2), as shown
below:
Standard
Specifications
CISPR 11
30 MHz – 1 GHz, Class A
CISPR 11
150 kHz – 30 MHz, Class A
IEC 61000‐3‐2
10 min, Class D (no limits for < 75W)
IEC 61000‐3‐3
120 min (no limits for < 75W)
IEC 61000‐4‐2
Contact: ± 2kV, ± 4kV, ± 8kV; Air: ± 2kV, ± 4kV, ± 8kV ± 15kV
IEC 61000‐4‐3
3 V/m, 80% AM at 1 kHz
80MHz – 2.7GHz
IEC 61000‐4‐4
Power Supply Lines, ± 2kV
100 kHz PRF
IEC 61000‐4‐5
Line to line ± 0.5kV, ± 1kV
IEC 61000‐4‐6
Power Supply Lines: 3 Vrms, 0.15 ‐ 80MHz;
ISM Bands
Input/Output Lines: 3 Vrms, 0.15 ‐ 80 MHz; 6 Vrms, ISM bands, 80% AM at 1 kHz
IEC 61000‐4‐8
30 A
/m, 50Hz or 60Hz
rms
IEC 61000‐4‐11
0 % U
: 250/300 cycles, 0°, 90°, 270°
T
10 - 33C (50 - 90F)
0 - 60C (32 - 140F)
5 - 95%, non-condensing
70 – 106 kPa
Dual scale: 5-60C and 40-140F
± 1.1°C (2°F)
F-270
F-300
1367 (361)
F-425
1828 (483)
23