Facility Requirements
The following are needed prior to using your Whirlpool:
Clean water source for filling tub. For burn patients with open wounds, follow the
physician's recommendation for water quality.
Drain plumbed per Figure 1, Figure 2, Figure 3, or Figure 4, installed by licensed
plumber
Medical-grade GFCI receptacle(s) located per Figure 1, Figure 2, Figure 3, or
Figure 4, installed by licensed electrician
Environmental conditions within those specified in the Specification section of this
manual
NOTE: The emissions characteristics of this equipment make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment
(for which CISPR 11 class B is normally required) this equipment might not offer adequate
protection to radio-frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the equipment.