At the other end of the DUAL electrode cable are two special connectors that plug into the
CELL socket on the back of the LC-4C detectors. The longer of these leads controls W2,
and the shorter controls W1. Make sure that the internal pins of these connectors are prop-
erly aligned before plugging in. The W1 lead can be used for SINGLE electrode applica-
tions, but the W2 lead CANNOT.
At the recorder, connect the leads as follows (see #6 above):
from W1 detector:
from W2 detector:
DO NOT use black jumper (green plug) to jumper LO to GND since this has already
been accomplished through the OUTPUT GNDs of the two LC-4Cs, as under NOTE
above. This prevents the occurrence of a ground loop (i.e., two grounds instead of
one).
white lead (red plug): "+" or "HI" recorder input
black lead (black plug): "–" or "LO" recorder input
black jumper (green plug): recorder ground terminal
(LO is thus jumpered to GND)
white lead (red plug): "+" or "HI" recorder input
black lead (black plug): "–" or "LO" recorder input