Cleaning and Disinfecting

Cover Care: For cover care instructions, refer to the manual provided with the cover.
Warnings:
- Cleaning and disinfecting are separate processes. Cleaning must precede disinfection. If the product becomes soiled, or between use by different individuals:
clean, disinfect, and check the product for proper functioning.
- Clean the product regularly and when it becomes soiled.
- Follow all manufacturer bleach container and germicidal disinfectant safety guidelines.
- Note about sterilization: High temperatures accelerate aging and will damage the product assembly. The products featured in this manual are not packaged
in a sterile condition, nor are they intended or required to be sterilized prior to use. If institution protocol demands sterilization: First, follow the cleaning and
disinfecting instructions; then, open the inflation valve and use the lowest possible sterilizing temperature, but not to exceed 70°C (158°F), for the shortest time
possible. DO NOT autoclave.
Cushion Care
Warning: The inflation valve MUST be closed to make sure that water does not enter the product during cleaning or disinfecting.
Cautions:
-
DO NOT machine wash or sanitize the cushion or machine dry the cushion.
-
Use of the following may cause damage to the cushion: abrasives (steel wool, scouring pads); caustic, automatic dishwasher detergents; cleaning products
that contain petroleum or organic solvents, including acetone, toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluid, adhesive remover; oil-based
lotions, lanolin, ozone gas; ultraviolet light cleaning methods.
-
DO NOT expose to direct sunlight.
Recommendation: When washing by hand, use room-temperature water.
Note: Deflated air cells may be easier to thoroughly clean. To deflate the air cells: with the inflation valve open, roll up and/or compress the cushion, and then
completely close the inflation valve.
To Clean the Cushion: Remove the cover. Close the inflation valve. Place in a large sink. Hand wash, using mild liquid hand soap, hand
dishwashing liquid, laundry detergent, or a multipurpose detergent (follow product label instructions). Gently scrub all cushion surfaces, using a
soft plastic bristle brush, a sponge or a cloth. Rinse with clean water. Air dry.
To Disinfect the Cushion: Follow the cleaning instructions, using 1 part household liquid bleach per 9 parts water. Keep the
product wet with the bleach solution for 10 minutes. Rinse thoroughly with clean water. Air dry.
Caution: Rinse thoroughly. Detergent residue may cause air cells to stick together.
Care Symbols
Hand wash.

Storage, Transport, Disposal and Recycling

Storage and Transport: Clean and disinfect the product before storing. Open the inflation valve(s). Deflate the product. Store the product in a container that
protects the product from moisture, contaminants and damage. After storing, follow instructions to prepare the product for use. After transport, inspect the
product to ensure it was not damaged. (see Troubleshooting).
Disposal: The components of the products in this manual are associated with no known environmental hazards when properly used and when disposed of
in accordance with all local/regional regulations. At the end of its useful life, treat as healthcare waste and dispose of according to local healthcare-waste
guidelines. Incinerate with other healthcare waste. Incineration must be performed by a qualified, licensed waste management facility.
Recycling: Contact a local recycling agency to determine recycling options for the product.

Limited Warranty

Limited Warranty Term from the date the product was originally purchased: cushion,12 months. The warranty does not apply to punctures, tears, burns, or misuse.
See also the instructions provided with the cover and the Limited Warranty supplement provided with your product, or contact Customer Support.
ROHO MOSAIC Cushion Operation Manual
1:9
Line dry in the shade.
5
Bleach (1 part bleach: 9 parts water).
+
1:9