Practical Considerations; Charging Physilog®5; Physilog®5 Configuration For Gait Measurements - Gait up PhysiGait Live User Manual And Outcome Parameters

Table of Contents

Advertisement

To display the report like in the mobile application, select this symbol. It opens a
new webpage where the report is displayed. To print the report or export to pdf,
select "Print" in the options of your browser or press "Ctrl + p".
Remark: on some browsers printing to pdf does not print all figures of the report.
For best results use Google Chrome and select "Print background images" in more
settings.
You can delete reports from the list if you want to remove them using the Bin symbol.
Attention, once you click on this symbol, the reports are deleted without further
notification or confirmation.

5. Practical Considerations

5.1
Charging Physilog®5
The Physilog®5 is charged via the microUSB port. Plug the microUSB connector of the cable
delivered with the Physilog®5 to the Physilog®. Plug the USB connector to a Computer or
Charger (voltage specifications: 5V, 100mA). Physilog®5 must be turned off for charging.
Do not charge Physilog® when room temperature is outside of range 0°C to 45°C.
Low battery level is indicated by orange blinking during measurement or if the Physilog®
blinks rapidly 3 times red when starting and doesn't start the recording. When the Physilog®
is discharged, plug it for charging for at least 30 minutes before use. An estimation of the
battery level of the Physilog®5 can be obtained when it is connected to the application.
5.2
Physilog®5 configuration for gait measurements
The configuration of Physilog®5 can be modified using Research ToolKit software or the
Android application. This allows to optimise the sensor settings for different applications. For
gait measurements the configuration must be set to the values indicated below, otherwise
the software cannot analyse the data. The correct configuration is pre-installed when you
order a PhysiGait Live, so you don't need to modify the configuration of the Physilog®5.
Sensor
Accelerometer
Gyroscope
Barometer
PhysiGait user manual v1.3.1 Date: 29.04.2019
Copyright © 2018, All Rights Reserved - Gait Up SA. CH-1015 Switzerland
Sampling rate
128Hz
128Hz
off
EPFL Innovation Park, Bat C. CH-1015 Lausanne
Sensor range
+/- 8 g
+/- 1000 deg/s
-
Gait Up SA
www.gaitup.com
TVA CHE-145.136.760
10

Advertisement

Table of Contents
loading

Table of Contents