Fluke i400s Calibration Information Manual page 6

Ac current clamp
Table of Contents

Advertisement

i400s
Calibration Information
i400s Range
1 mV/A
1 mV/A
1 mV/A
1 mV/A
10 mV/A
10 mV/A
10 mV/A
10 mV/A
6
1.800.561.8187
Table 3. Accuracy Tests
5520A Output
Nominal Value
LCOMP = ON
20 A, 50 Hz
0.4 A, 50 Hz
20 A, 400 Hz
0.4 A, 400 Hz
400 A, 50 Hz
8.0 A, 50 Hz
400 A, 100 Hz
8.0 A, 100 Hz
2 A, 50 Hz
40 mA, 50 Hz
2 A, 400 Hz
40 mA, 400 Hz
40 A, 50 Hz
0.8 A, 50 Hz
40 A, 100 Hz
0.8 A, 100 Hz
Figure 2. Current Accuracy Test Connections
www.
Multimeter Reading Limits
Lower
19.56
19.56
391.96
391.96
19.45
19.45
391.85
391.85
information@itm.com
.com
Upper
20.44
20.44
408.04
408.04
20.55
20.55
408.15
408.15
bbc01f.eps