3.4. Prior to Operation

3.4.1. Precautions and Limitations of Use

Check that,
There is no sample that can harm the chamber surface when heated,
Liquids that may be dilate or samples that may be melt are not heated in sealed containers,
The boiling point of the samples are higher than the set temperature,
The freezing points of the samples are lower than the set the temperature,
The liquids which may expand during heating do not overflow from their containers,
The set temperature does not destroy the structure of the samples,
The vapour and gases which are generated during the operation are not harmful to human
health or flammable or explosive,
Make sure that the safety thermostat is adjusted to the temperatures which are higher than
the set temperature,
The samples are not put on the chamber floor but on the shelves.

3.4.2. Water Filling

For the humidity functions DISTILLED WATER should be used in order to protect the
materials and the heaters of the instrument. The water filling should make from water filling
inlet (see figure) or the optional automatic water supply system can be used.