Above window, which allows adjustment of 5 pieces of the lamp at the door and have been
symbolized by the "TIME 1" and "TIME 2", is opened. "TIME 1" and "TIME 2"; At
different times during the day to arrange the lighting to be active provides the starting and
ending time periods.
By using "TAB" key and "VALUE INCREASE-DECREASE" keys the start and stop periods
are set as a daily period that uses the real time clock of the instrument and "ENTER" key is
pressed to save the settings for each page. 'ON' is chosen to light on the lamps, 'OFF' is
chosen to light off the lamps.
The same adjustments are performed for the "TIME 2".
The illumination can be activated when the door open. 'ON' is chosen to light on the lamps,
'OFF' chosen to light off the lamps.
After programming, the illumination can be activated by using the "F1" key below 'START'
expression and can be quitted from the illumination setting menu by using the "F4" key below
the 'EXIT' expression.
The program and illumination can be performed separately.
WARNING: If the illumination is at "START" position, date and time can not be changed.

4. 5. Completion of the Operation

See that the program is over.
If the printer was used during the run and if there is no need to print stand-by values stop
printing. (see 4.4.4)
Take the samples out. Be careful while handling the samples after the operation as they
can be very hot or cold.
Wipe the chamber surface if needed while the chamber is at room temperature.
You may leave the equipment at the stand-by position or switch it off.
WARNING: DO NOT OPEN THE DOOR WHEN THE TEMPERATURE AND
HUMIDITY LEVELS ARE HIGH INSIDE THE CHAMBER. THIS MAY CAUSE
BURNS ON HANDS AND FACE.