Summary of Contents for Dolomite MITOS P-PUMP REMOTE CHAMBER 400

  • Page 1 MITOS P-PUMP REMOTE CHAMBER 400 USER INSTRUCTIONS...
  • Page 2: Table Of Contents

    Contents Mitos 400 Safety Information System Set-up Introduction Using the Mitos P-Pump Remote Chamber Mitos P-Pump Control Control SHPT-487168127-356_v.1.0 Page 2 of 12...
  • Page 3 Mitos P-Pump product range Product name Product picture Product description The Mitos P-Pump provides pulseless liquid flow with a precise pressure driven pumping Mitos P-Pump mechanism. It operates over a wide pressure (Part No. 3200016) range (0-10bar) with excellent response time and accuracy.
  • Page 4 The Mitos P-Pump 3-way Chamber Lid can be Mitos P-Pump used to pump 3 liquids simultaneously from the 3-way Chamber Lid Mitos P-Pump. This provides a low cost method (Part No. 3200044) of delivering 3 reagents to a microfluidic system with pulseless pressure pump technology.
  • Page 5: Mitos 400 Safety Information

    This may result in failure of the Remote Chamber and injury to the user. Do not use without the safety screen in place. Please contact Dolomite for replacement of the glass vessel. Maximum Allowable Temperature Range 0-50°C Do not subject the remote chamber to internal or external temperatures over 50°C or below 0°C.
  • Page 6 Use of Oxygen and flammable gases The remote chamber should only be used with class 2 fluids. The user is responsible for ensuring the chamber is suitable for use with the intended fluids. The chamber should not be used with Oxygen or flammable gases. Compatible fluids/gases User must ensure that all gasses and fluids used with the remote chamber are compatible with the materials used in the construction of the...
  • Page 7: System Set-Up

    2 System Set-up 2.1 Introduction atmospheric pressure liquid flow flow gas flow Mitos P-Pump chamber pressure is equal to Remote Chamber pressure supply The Mitos P-Pump Remote Chamber is connected to the Mitos P-pump, which precisely controls the pressure inside both chambers. In this mode of operation, the liquid vessel is placed inside the Mitos P-Pump Remote Chamber.
  • Page 8: Using The Mitos P-Pump Remote Chamber

    2.2 Using the Mitos P-Pump Remote Chamber Check that you have the following components: Duran bottle • Remote chamber • Remote chamber lid with dip tube fitting • Pneumatic connector • 250ml Duran bottle remote chamber lid remote chamber • Mitos P-Pump (not included) •...
  • Page 9 Step 4: Place liquid vessel inside remote chamber. DO NOT PLACE LID ON LIQUID VESSEL This is important to prevent pressure chamber failure and injury to the user. Step 5: Place lid on remote chamber and tighten the 4 thumb screws evenly. Do not completely tighten one thumb screw before tightening the others as this will result in uneven clamping...
  • Page 10 Step 7: Use the pneumatic connector supplied to connect the Remote Chamber to the Mitos P-Pump. Step 8: Set pressure on Mitos P-Pump (see Mitos P-Pump user instructions) to start pumping. To stop pumping, vent the chamber by pressing the back button on the Mitos P-Pump.
  • Page 11: Mitos P-Pump Control

    3 Mitos P-Pump Control 3.1 Control When operating with the Mitos P-Pump Remote Chamber 400, it will take significantly more time for the chamber to reach the desired pressure due to the increased volume. However, the Mitos P-Pump will still be able to achieve very stable control of the pressure.
  • Page 12 Dolomite Microfluidics 3 Anglian Business Park, Royston, SG8 5TW, UK T: +44 (0)1763 242491 E: info@dolomite-microfluidics.com W: dolomite-microfluidics.com SHPT-487168127-356_v.1.0 Page 12 of 12...