Reprocessing; Single Patient Use - Pari COMPACT2 Instructions For Use Manual

Inhalation system
Hide thumbs Also See for COMPACT2:
Table of Contents

Advertisement

4

REPROCESSING

DANGER
Danger of electrocution
Liquids can conduct electricity, thereby posing a risk of elec-
tric shock.
• Always switch the compressor off and disconnect the power
plug from the mains socket before starting to clean it.
If the inhalation system is used by only one patient, follow the
instructions for cleaning and disinfecting in the section "Single
patient use" [see: Single patient use, page 26].
If the inhalation system is used by several patients, follow the
instructions for cleaning and disinfecting in the section "Multiple
patient use" [see: Multiple patient use, page 30].
If the inhalation system is used in professional environ-
ments, follow the information on reprocessing included in the
appendix at the end of these instructions for use.

4.1 Single patient use

Cleaning and disinfection cycles
Compressor housing
Nebuliser (without con-
nection tubing) and mask
Air filter
Nebuliser and mask
The mask can be cleaned, disinfected and dried together with
the nebuliser.
Preparation
MASK
• Detach the mask from the nebuliser.
• Detach the elastic band from the mask.
– 26 –
Clean when visibly dirty
– Clean immediately after every use
– Disinfect at least once a day
Replace after 200 operating hours
(approx. 1 year)