Reprocessing In Home Environments; Reprocessing Cycles; Processing Limits - Pari PEP S System Instructions For Use Manual

Hide thumbs Also See for PEP S System:
Table of Contents

Advertisement

4
REPROCESSING IN HOME
ENVIRONMENTS
The product components must be cleaned thoroughly immedi-
ately after each application and disinfected at least once a
week.
The connection tubing cannot be cleaned or disinfected.
Dry the connection tubing after each use [see: Care of the
pressure gauge and connection tubing, page 24].
The maximum operating life of the connection tubing is 1 year.

4.1 Reprocessing cycles

PARI PEP S System
Pressure gauge

4.2 Processing limits

PARI PEP S System,
disinfection
– 19 –
– Clean immediately after every use
– Disinfect once per week
Cleaning and disinfection can be carried
out together with the nebuliser.
Care as required [see: Care of the
pressure gauge and connection tubing,
page 24].
300 processing cycles,
max. 1 year