Multiple Patient Use - Pari COMPACT2 Instructions For Use Manual

Inhalation system
Hide thumbs Also See for COMPACT2:
Table of Contents

Advertisement

Clean the compressor housing as follows:
• Wipe the outer surface of the
housing with a clean, damp
cloth.

4.2 Multiple patient use

Cleaning and disinfection cycles
Nebuliser (with
connection tubing) and
mask
Compressor housing
Air filter
Nebuliser and mask
The nebuliser and mask are not suitable for use by multiple
patients. Use a separate nebuliser (including connection
tubing) and a separate mask for each patient.
Disinfecting the compressor housing
For disinfection, use a standard, alcohol-based disinfectant
(e.g., isopropanol). Regarding the use and quantity of the disin-
fectant, it is imperative to follow the information for use of the
agent used.
To disinfect the compressor, proceed as follows:
– 30 –
Replace before each patient change
Disinfect before each patient change
Replace after 200 operating hours
(approx. 1 year)