  Blanket tracker 
 
 
 
 
 
  
 
C83D5EN12 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Advertisement

Table of Contents
loading

Summary of Contents for Cerlic Blanket Tracker

  • Page 1 Blanket        Tracker                                   Blanket tracker                  2011‐11‐09    C83D5EN12 ...
  • Page 2: Table Of Contents

    Blanket        Tracker    Table of Content      TABLE OF CONTENT ............................  2    ..............................3   ENERAL INFORMATION   ..............................3   ECURITY INFORMATION   .............................3   ESCRIPTION OF THE FUNCTION   ............................3   EASUREMENT  UNCTIONS  ..................................3   ENSOR BODY UNPACKING ................................  4   Cable winder  .................................. 4   HANDLING AND INTERFACE ...
  • Page 3: General Information

    Blanket        Tracker    General information  It is of great importance to read all parts of this manual prior to start up of the  instrument. If the Blanket Tracker is not used and handled according to this manual,  then the life and functionality may be jeopardized, and all warranties will be void.     Security information  This instrument should be handled by qualified and authorized personnel only.  It is mandatory to follow all security and other routines that apply at site when  using the Blanket Tracker in tanks and basins.     Within EU it is prohibited to dispose of electric and electronic waste in regular  waste as these may contain harmful substances. All electric and electronic waste  must be sorted and left for recycling. Such products are labeled with an X‐marked  waste bin. It is important that everyone cooperate in order to secure a high level when it  comes to recycling and help to save our environment. If such waste not is handled and  recycled according to regulation (EC Directive 2002/96/EC) the environment as well as  people’s health may be jeopardized.    Description of the function  Blanket Tracker is a portable optical suspended solids meter designed to measure  sludge blanket depth and suspended solids in clarifiers, thickeners, aeration basins, etc.  in water and wastewater plants, as well in other facilities. It is possible to display  measured values as text or as a graphic image of the sludge profile. Two different alarm  levels for the sludge blanket may be set to indicate fluff and sludge blanket levels. These  may be displayed next to the graphic sludge profile.    Measurement Functions  The Blanket Tracker is able to store up to 250 measurements as a graphic profile, each  with information about time, date, location, that may be named by the user (maximum  10 characters e.g. Clarifier 2; Thickener 5, etc.) Each profile also contains data about  blanket and fluff depths that applied for the specific measurement.    Sensor body  The sensor body contains optics and electronics that should not be exposed to  mechanical abuse or high temperatures. If the sensor body has mechanical damages, ...
  • Page 4: Unpacking

    Blanket        Tracker      Unpacking   Open the instrument case and check that no damages have occurred during shipping.  The instrument case should contain the following items plus the service manual, see  figure 1.      fig 1    Cable winder  In order to facilitate the handling of the Blanket Tracker it is supplied with a SS cable  winder that is attached to the battery case, see fig 2. It is possible to wind up part or all  of 33’ cable on the winder.. The cable is locked in place with an O‐ring, see fig 3.                                                                                                        fig 2                                                               fig 3      4   ...
  • Page 5: Handling And Interface

    Blanket        Tracker      Handling and Interface   The following icons show the keypad push buttons and describe their functions.     Opens the main menu or confirms (enters) a choice in the menus.      Closes a menu choice without changing, or exit’s  one‐step backwards  in the menu tree (escape function).     Navigates up one step in the menu tree or increases the value of a  chosen number in a menu.     Navigates down one step in the menu tree or decreases the value of a  chosen number in a menu.     The profile button alternates between measuring values in text or  sludge profile presentation.     Restart of a measurement. At restart, a new measuring will be initiated  and replaces the existing value. The measured values may be stored at an  optional location in the log file.     Turns on the backlight for display for two minutes.     On/Off switch. Press the button in order to turn on or off the Blanket  Tracker.                    ...
  • Page 6: Description Of The Display In Text Mode

    Blanket        Tracker      Description of the Display in Text Mode  At start up, the display is shown in fig 4 and description in fig 5. Value for solids  concentration and depth are displayed on the left and alarm values for fluff (6) and  sludge blanket (7) with actual depth for each as (9) & (10) below.  Depths are shown as  Blind until alarm for each concentration goes off.  Plus date, time and temperature  across the top.                                                                       ...
  • Page 7: Description Of The Display In Profile Mode

    Blanket        Tracker      Description of the Display in Profile Mode  When pressing the Profile button marked   the display will show the sludge profile.  See fig 6 below. At any time during a measurement it is possible to switch between text  mode and profile mode by pressing the button   without affecting the measuring  value. General information such as sludge concentration and depth is always shown in  either the text or profile modes windows.                                                 ...
  • Page 8: Main Menu

    Blanket        Tracker      Main menu  Press   to open the main menu. See fig 8.  Use   or   to pick the desired sub menu and open with  .         fig 8  Settings   In the SETTINGS menu (fig 9) it is possible to set  alarm values for fluff and sludge  blanket concentrations. In this menu the span (max depth) measurement is set to fit the  profile scale maximum depth. It is recommended to set a ”blind zone” so that the solids  concentration measurement reading start just below the water surface like 1” or so    This is done to avoid optical light reflections that could otherwise cause false alarm of  fluff and blanket depths. Measure Mode allows you to pick DEPTH for a normal clarifier  and HEIGHT for an upside down clarifier like a DAF so you are measuring from the  surface down for blanket.         fig 9      8   ...
  • Page 9: Threshold 1  (Fluff)

    Blanket        Tracker      Threshold 1  (Fluff)  This value defines the solids concentration that indicates the FLUFF level. The  concentration may be given in g/l, mg/l, % or ppm For more information see ”Settings”.  When the preset alarm value is reached, then this will be shown on the display as FLUFF  LEVEL. The sensitivity for fluff concentration is normally set at 10% to 25%  of the  sludge blanket alarm  value.      fig 10    Press   to open the menu and select ”SETTINGS” confirm with  . Step  down with   and pick ”Threshold 1” with  . Step up with   or down  with   until desired concentration is obtained. Confirm and step to next  digit with  . Exit the menu with   until one of the two     main windows appear (text mode or profile mode).            9   ...
  • Page 10: Threshold  2 (Sludge Blanket)

    Blanket        Tracker      Threshold  2 (Sludge Blanket)  This value defines the solids concentration that defines the SLUDGE BLANKET Level.  The concentration may be given in g/l, mg/l, % or ppm For more information see  ”Settings”. When the preset  alarm value is reached, then this will be shown on the  display as SLUDGE BLANKET level. In a clarifier this concentration is normally choosen  to correspond with the return sludge (RAS) concentration, e.g. 5,000 mg/l (ppm) and in  a thickener e.g. 7,000 mg/l.      fig 11    Press   to open the menu and select ”SETTINGS” confirm with  . Step  down with   and select ”Thrreshold 2 ” with  . Step up with   or  down with   until desired concentration  is obtained. Confirm and step to  next digit with  . Exit the menu with       until one of the two main windos appear (text mode or profile mode).  Maximum depth  Press   to open the menu and select “SETTINGS”. Confirm with  . Step  down with   and select ”MAX DEPTH” with  . To scale the “MAX  DEPTH” span that is shown at profile measurement, step up with   or ...
  • Page 11: Measuring Mode

    Blanket        Tracker      Measuring mode  In order to define how the levels for FLUFF and SLUDGE BLANKET are presented, it is  possible to select if the depth is shown from the surface down (Depth) or from the  bottom and up(Height) for a DAF. To choose between these options perform the  following steps;    Press   to open the menu and select ”SETTINGS”. Confirm with  . Step  down with   and select ”MEAS. MODE” with  . Step up with   or  down with   in order to select desired mode. Confirm with  . Exit the  menu with   until one of the two main windos appear (text mode or  profile     mode).  SETUP  The SETUP menu (fig 12) enables selection of language, units for depth and  concentration units, temperature units, time and date.       fig 12  Language  The following languages are available; Swedish, English, German and French. To select  the desired language, perform the following steps;    Press   to open the menu and step down with   and select ”SETUP” with ...
  • Page 12: Depth Unit

    Blanket        Tracker      Depth Unit  Following units for depth measurement are available; cm, dm, m, inch and ft (with  inches).    Press   to open the menu and step down with   Select.”SETUP” with  . Step down with   and select ”Depth unit”. Confirm with  . Step up  with   or down with   until desired unit (cm, dm, m, in, ft). Confirm  with  . When all settings are done, exit with   until one of the two main  windos appear (text mode or profile     mode).    Concentration Units  Following units for sludge concentration are available; g/l, mg/l, %, and ppm.    Press   to open the menu and step down with  . Select ”SETUP” with  . Step down with   and selsct ”Conc  unit”. Confirm with  . Step  upwith   or down with   until desired unit (g/l, mg/l, %, ppm). Confirm  with  . When all settings are done, exit with ...
  • Page 13: Time And Date

    Blanket        Tracker      Time and date  To set time and date;  Press   to open the menu and step down with  . Select ”SETUP” with  . Step down with   and select ”Time and Date”. Confirm with  . Step  up with   or down with   until desired time and date format. Step up  with   or down with   and set actual time and date. Confirm and step  further with  . When all settings are done, exit with   until one of the  two main windows appear (text mode or profile mode).          fig 13          13   ...
  • Page 14: Calibration

    Blanket        Tracker      Calibration  Calibration can be done for both depth measurement and sludge concentration. For  depth measurement, zero compensation is performed every time when the Blanket  Tracker is turned on to adjust for barometric pressure. It is also possible to perform a  manual calibration of zero and one meter (3.3 ft) depth. This is done by starting the  Blanket Tracker and perform a zero calibration of the depth.          fig 14      Zero Sample   Press   to open the menu and step down with  . Select ”Calibration”  with  . Step down with   and select ”zero sample”. Confirm with  .  Put the sensor in clean, deareated water and confirm with  .  . When  zero calibration is done, exit with   until one of the two main windows  appears (text mode or profile mode).    Conc. sample – Sludge Sample   Press   to open the menu and step down with  . Select ”Calibration”  with  . Step down with  and select ”Conc sample”. Confirm with ...
  • Page 15: Lab Value

    Blanket        Tracker      Lab Value   Press   to open the menu and step down with  . Select ”Calibration”  with  . Step down with   and select ”lab value ”. Confirm with  . Step  up with   or down with   until the value corresponds with the lab   value.  Confirm with  . When all settings are done, exit with   until one  of the two main windows appears (text mode or profile mode).    0 m Cal. ‐ Zero depth calibration  Press   to open the menu and step down with  . Select ”Calibration”  with  . Step down with   and select ”0 m cal. ”. Hold sensor vertically in  the air and start zero calibratoin  with  . Continue until value is stable and  then save calibration with  . Exit the menu with .    1 m Cal. – 1 m or 39.4” depth calibration  Press   to open the menu and step down with  . Select ”Calibration”  with ...
  • Page 16: Log

    Blanket        Tracker      Log  Up to 250 measurements may be stored in the Blanket Tracker internal memory. Every  stored measurement is saved as a profile that contains all data available at the actual  measurement. Each individual line in the log may be tagged with 10 alphanumerical  characters; e.g. Clarifier 2; Thickener 5, etc.       fig 15      Press   to open the menu and step down with  . Select ”Log” and  confirm with  . Step up with   or down with   to pick the right log  datas. Confirm with  . When all settings are done, exit with   until one  of the two main windows appear (text mode or profile mode).        fig 16    Step down with   and select Position ‐  Show, Empty, Delete, or Create”.  Confirm with  . ”Delete” and ”Empty” are local and are only valid in its  choosen position. ”Show” or ”Create” are global and valid for all positions.  Step up with   or down with   to navigate the menu tree.Confirm with ...
  • Page 17: Getting Started

    Blanket        Tracker    Getting Started  Getting started with the Blanket Tracker  Start the Blanket Tracker by pressing the button marked  . To switch OFF/ON the  unit, press the same button. The unit will do an atmospheric pressure compensation at  start‐up and measurement will start automatically.  Values for sludge concentration and  depth are displayed in clear text and real time. If the unit is not active during an eight  minute period then it will automatically be turn off without saving the measurements.    In order to get correct scaling of the depth range in profile range shown on the display,  then it is necessary to set the maximal depth at the actual measuring position.    Setting of Maximum depth  Press   to open the menu and select ”Settings” with  . Step down with   until  desired Max Depth is selected with  .      fig 17      Select the depth value numbers by stepping   and change the values with   or  .  Save by pressing   to save.        fig 18   ...
  • Page 18: Profile Mode

    Blanket        Tracker      Profile mode  In order to switch between text mode and profile mode, press the button marked  . It  is possible to step back and forth with this button without affecting the actual  measurement.    Save a profile  To save a measurement profile, or to perform a new measurement, press  , When  doing this, an option to save  or restart the measurement   will be presented,  There is a short command from the text mode display where it is possible to select  position for storing the measurement profile. Step up with   or down with   to  reach desired log position. The postion tag will be shown in the lower margin in the  display window. Fig 19 below, shows that the actual profile will be stored at position   ”SED_1”          Fig 19      It is possible to tag each of the 250 log positions with up to 10 alphanumerical  characters. By entering the log position by tag name, all actual information for the stored  measurement may be retrieved.    Alarm  values  At delivery the Blanket Tracker is preset with the following values; Fluff (1000mg/l) and  Sludge Blanket (5000mg/l). Using these values, it is possible to instantly start to  measure and get a picture of the sludge profile. To change the settings please see ...
  • Page 19: Maintenance

      In order to verify if a depth calibration is needed, lower the sensor into water until the  one meter cable mark. The depth value should not differ more than ±2cm (±1”). A higher  deviation requires a zero and one meter calibration, see section Calibration.    When cleaning the sensor, it is preferred to use the Cerlic Sensor Cleaning (CSC) liquid.  It is also possible to use a soft cloth and water. Pay attention not to scratch the  measuring windows or damage the built in pressure sensor.  Changing of Batteries  Cerlic Blanket Tracker is supplied with four AA batteries, placed inside the unit. To change the batteries, follow the steps below; 1. Open the back part of the handle on the enclosure. This part is a  combined battery lid and cable winder. This is done by removing  the two  screws countersunk into the back of the enclosure, see fig 20. Do not  remove the screw holding the cable winder.    2. Remove the battery lid and replace the batteries. Make sure to follow the  label showing the polarity (+/‐).    3. Remount the battery lid. ...
  • Page 20: Trouble Shooting

    Blanket        Tracker      Trouble shooting  Check that the batteries are in good condition. The Blanket Tracker has a built in logic to  increase the battery life. If the battery voltage goes  below a certain level, then the  software will block the following functions; display back light, acoustic and vibration  signals.    If the Blanket Tracker values for depth and concentration should deviate, a new  calibration for depth and/or concentration. Please see section Calibration for  instructions.    In case of any malfunction that is not possible to correct with a new calibration, please  contact Cerlic or a local Cerlic representative. In case of sending the Blanket Tracker to  Cerlic for check up or repair, please make sure to use the form for Return of Material  (RMA) prior to dispatch. The RMA document can be downloaded from the Cerlic web  page www.cerlic.com. The WEB page also has the actual and correct receiving address.   Spares    The Blanket Tracker is supplied with four AA batteries. The batteries are the only parts  that are subject to be changed by the user. It is recommended to keep an extra set of  batteries available.    Spare part list:      Part no Description     21450731 Battery 3V 200mAh   21450989 Battery AA  ...
  • Page 21: Contact Information

    Blanket        Tracker      Contact information    Actual visit and receiving address is always available at the Cerlic web page.     Europe  Cerlic Controls AB  Mälarvägen 3, SE 141 71 SEGELTORP, Sweden  Phone:+46 850 169 400/Fax: +46 850 169 429  Mail address: P.O. Box 5084, SE‐141 05 KUNGENS KURVA, Sweden   web: www.cerlic.com  US & Canada  Cerlic Enviromental Controls, Inc  200 Burdette Road  Atlanta, GA 30327  Phone: 404‐256‐3097  Warranty   Instruments delivered from Cerlic Controls AB, are carefully checked and tested prior to  the shipment.     1.  Cerlic warranty and repair  free of charge the Portable Blanket Tracker if it is  determined that the problem is  any fault in manufacturing or equipment during  the warranty period.    2.  The warranty period is 12 months from the date of invoice.    21   ...
  • Page 22: Specification, Portable Blanket Tracker

    Blanket        Tracker    Specification, Portable Blanket Tracker   Function   Sludge Concentration and Depth  Measuring principle  Optical light transmission  Wave length   NIR 850 nm  Measuring range  Max –  ~10 000 mg/l (ppm)  Accuracy suspended solids  1% FS (full scale)   Repeatability   < 2% of Measured Value  Type of Measurement   Continuous with Profile  Number of Detectable Levels  Two (fluff and sludge blanket)  Measuring units   g/l, mg/l, %, ppm  Depth units  cm, dm, m, in, ft with inches  Principe of Depth Measurement  Pressure cell, absolute pressure  Accuracy Depth Measurement  +/‐ 0,5% FS   Maximum depth   19m (62 ft)  Display  Graphical, 128  x 64 pixels, LCD   Back light  Yes  Languages  Swedish, English, German, French  Log Function   250 Measurements with profile ...