LEEC GA2000 Operating Instructions Manual page 3

Automatic co2 incubators
Table of Contents

Advertisement

1.8
FUSES
A 6.13A fuse next to the power cable entry point limits the current that the incubator
can draw. There are additional internal fuses located under the removable lid for
the main chamber heaters, door heater and decontamination system (GA2000 &
GA3000 models only).
1.9
ALARMS
The LEEC multifunction controller incorporates a comprehensive alarm system.
There are over and under alarms for both temperature and CO2, which activate a
buzzer and start the displays flashing. Your incubator is also fitted with a remote
alarm socket which provides volt-free change over contacts for switching
any remote alarm (e.g. Building Management System) that you may connect to it.
All models are fitted with an independent over heat safety cut out which is
independent of the main controller has a separate sensor.
3

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GA2000 and is the answer not in the manual?

This manual is also suitable for:

Ga2010Ga3000Ga3010

Table of Contents