Summary of Contents for LEEC C100

  • Page 1: Table Of Contents

    USER MANUAL LEEC CLASSIC INCUBATOR C100, C157 and C330 CONTENTS GENERAL DESCRIPTION INSTALLATION AND SWITCHING ON INCUBATOR PROGRAMMING TEMPERATURE OVER TEMPERATURE CUT OUT CLEANING & DISINFECTING WARRANTY - PREVENTATIVE MAINTENANCE V1.0 March 2009 LEEC Limited Private Road No. 7 Colwick Industrial Estate...
  • Page 2: General Description

    TEMPERATURE CONTROLLER A very easy to program electronic PID temperature controller is fitted to the LEEC Classic Incubator range. This controller is easy to use and enables quick programming of the incubator chamber temperature.
  • Page 3 * Please refer to your employers/organisation Safe Manual Handling Policy: The Manual Handling Operations Regulations 1992, before attempting to lift the incubators. The weights are as follows: C100 – 54kg, C157 – 70kg and C330 – 107kg. CONNECTION TO THE ELECTRICAL SUPPLY Your LEEC incubator is supplied with a 2 metre mains cable.
  • Page 4 To switch on the incubator press in the Power button on the front facia panel. This button will illuminate Green when power is On. Front Facia Panel Power On The incubator display will now light up displaying both the programmed temperature (SET) and actual inner chamber temperature (ACT) and the Heater (H) LED on the display indicates when the heaters are active.
  • Page 5: Programming Temperature

    PROGRAMMING TEMPERATURE DISPLAY The programmed temperature, Set Point (SET) is displayed in Green and the Actual Chamber Temperature inside the Incubator Inner Chamber (ACT) is displayed in LED Indicator next to the letter H (Heaters) on the display illuminates when the heaters are on.
  • Page 6: Over Temperature Cut Out

    3 seconds. If the Low / High alarm values or the 2 minutes dwell need to be adjusted then please contact LEEC Limited for instructions. INDEPENDENT OVER TEMPERATURE SAFETY CUT-OUT An independent fail safe over temperature safety cut out is fitted to this incubator. The over temperature cut out is factory set to 40°...
  • Page 7 RESETTING OVER TEMPERATURE CUT-OUT In the unlikely event of the over temperature safety cut out activating, this can be reset by Pressing in the Button at the Back of the incubator. Over Temperature Cut Out at Rear of Incubator Push In Red Button to Reset PROCEDURE FOR PROGRAMMING OVER TEMPERATURE CUT OUT IF OPERATING TEMPERATURE IS ABOVE 37.0°...
  • Page 8: Cleaning & Disinfecting

    However, corrosion can still result from improper use of fungicides and bactericides. Never use any Chloride based chemicals for cleaning as this can result in permanent damage to the stainless inner chamber and this is not covered by LEEC warranty as no stainless steel is completely resistant to chlorine.