LEEC GA2000 Operating Instructions Manual

Automatic co2 incubators

Advertisement

OPERATING INSTRUCTIONS FOR
LEEC AUTOMATIC CO2 INCUBATORS
MODELS GA2000 and GA2010
MODELS GA3000 and GA3010
With Infrared CO2 detectors
1.0
2.0
3.0
4.0
5.0
6.0
7.0
8.0
9.0
10.0
11.0
12.0
DIAGRAMS AND SPECIFICATION
CONTENTS
1
Revised: April 2009
LEEC Limited
Private Road No. 7
Colwick Industrial Estate
Nottingham
NG4 2AJ
Tel. 0115 961 6222
Fax. 0115 961 6680
Web www.leec.co.uk
E-mail sales@leec.co.uk

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GA2000 and is the answer not in the manual?

Questions and answers

Summary of Contents for LEEC GA2000

  • Page 1: Table Of Contents

    OPERATING INSTRUCTIONS FOR LEEC AUTOMATIC CO2 INCUBATORS MODELS GA2000 and GA2010 MODELS GA3000 and GA3010 With Infrared CO2 detectors CONTENTS GENERAL DESCRIPTION INSTALLATION AND SETUP DESCRIPTION OF CONTROLS OPERATING PROCEDURE FOR NORMAL RUNNING SETTING OPERATING PARAMETERS CHECKING TEMPERATURE CALIBRATION CHECKING CO2 CALIBRATION...
  • Page 2: General Description

    This keeps the view into the chamber clear. COOLING A cooling coil is built in to your incubator which, when connected to a LEEC Self Contained Recirculating Cooler, allows the incubator to operate at temperatures near or below ambient.
  • Page 3 A 6.13A fuse next to the power cable entry point limits the current that the incubator can draw. There are additional internal fuses located under the removable lid for the main chamber heaters, door heater and decontamination system (GA2000 & GA3000 models only).
  • Page 4: Installation And Setup

    Your LEEC CO2 incubator must be installed in suitable location. It should be placed on a firm flat surface such as a laboratory bench, LEEC stacking frame or the floor. All models have adjustable feet at the front. An adjustable spanner is required if the feet need adjusting.
  • Page 5: Description Of Controls

    CO2 level without controlling the gas. Note: The outer door must be closed for the CO2 supply solenoids to function. DECONTAMINATE KEY (models GA2000 & GA3000 only) This key switch on the control panel initiates decontamination (see section 11.0).
  • Page 6: Operating Procedure For Normal Running

    Door LED This amber LED shows when the door heater is working. It flashes during normal operation. 3.12 Decon. LED (models GA2000 & GA3000 only) This amber LED shows when the decontamination cycle is active. 3.13 % CO2 DISPLAY This displays the chamber %CO2 reading during normal running. It is also used in the setting of various parameters (see section 5.0).
  • Page 7: Setting Operating Parameters

    SETTING USER OPERATING PARAMETERS The temperature and CO2 concentration inside the chamber can be set by the user. In addition, various alarm parameters can also be adjusted. This section describes the functions and the procedures and how to change them. There is also an Engineering Mode which can only accessed by trained service personnel.
  • Page 8 Temperature Set Point Displays Command Action Temp Remarks Display Press ° C Set SP_t 37.0 Unit factory set to temperature set button once 37.0 point Press ▼ SP_t xx.x Value decreases Select temperature set Value increases Press ▲ point SP_t xx.x (where xx.x is new value)
  • Page 9 Set Temperature Alarm Delay Time Displays Command Action Temp Remarks Display Press ° C Set tdEL 12.0 tdEL sets delay temperature button 4 times for temperature alarm delay low alarm. time There is no delay on the high temperature alarm Press ▼...
  • Page 10 Set Temperature Offset Display Command Action Temp Remarks Display Press ° C Set oFFS Adds offset to temperature button 6 times temperature offset display reading Press ▼ oFFS xx.x Value Select decreases temperature offset Value increases Press ▲ oFFS xx.x (where xx.x is new value) Press ◄╝...
  • Page 11 Set CO2 Gas Operating Level Displays Command Action Temp Remarks Display CO2 Press CO2 Set SP_c Sets % CO2 level set point button once operating level. CO2 level is factory set to 0.0% to disable all CO2 alarms Press ▼ SP_c xx.x Value decreases...
  • Page 12 5.10 Set CO2 Gas Alarm Delay Time Displays Command Action Temp Remarks Display CO2 Press CO2 Set cdEL 12.0 Sets delay time alarm delay button 4 times for both high and time low %CO2 alarms Press ▼ cdEL xx.x Value decreases Select CO2 alarm delay Value increases...
  • Page 13: Checking Temperature Calibration

    COOLING COIL A cooling coil is built in to your incubator which, when connected to a LEEC Self Contained Recirculating Cooler, allows the incubator to operate at temperatures near or below ambient. The coil is located near the fan at the top of the chamber behind a false panel to give maximum efficiency.
  • Page 14: Maintenance And Cleaning

    CO2 gas is injected into the chamber. 10.4 CHAMBER CORROSION Inner chambers of LEEC CO2 incubators are made from the finest quality stainless steel available. However, corrosion can still result from the improper use of fungicides and bactericides. Chlorine based chemicals used for cleaning can cause permanent damage to the chamber and this is not covered by the warranty as no stainless steel is completely resistant to chlorine.
  • Page 15: Decontamination Procedure

    11.0 HIGH TEMPERATURE DECONTAMINATION (Research models only) Models GA2000 and GA3000 are fitted with a high temperature decontamination facility as standard. An additional heating element raises the chamber air temperature to approximately 95° C to decontaminate the chamber. This can be performed when required.
  • Page 16 GA2000, GA2010, GA3000, GA3010 FACTORY DEFAULT PARAMETERS Temperature Value SP_t 37.0 lo_t -1.0 hi_t tdEL 12.0 40.0 oFFS Value SP_c lo_c -1.0 hi_c cdEL 12.0 oFFS...