METAL X SYSTEM FACILITIES GUIDE TABLE OF CONTENTS OVERVIEW INSTALLATION PLANNING SHIPPING AND UNLOADING SYSTEM COMPOSITION MOVING INFORMATION FACILITIES PREPARATION OPERATIONAL SPACE PLANNING WASTE DISPOSAL ACCESS REQUIREMENTS ELECTRICAL ACCESS ADDITIONAL INFORMATION VENTILATION AND AIRFLOW NOTE ON CONNECTIVITY TECHNICAL SUPPORT LOCAL SOLVENT REGULATIONS...
Page 4
METAL X SYSTEM FACILITIES GUIDE TABLE OF CONTENTS SINTER-2 ELECTRICAL SPECIFICATIONS GROUND FAULT PROTECTION OVERCURRENT PROTECTION DEDICATED DISCONNECT ENVIRONMENTAL REQUIREMENTS CLIMATE CONTROL VENTILATION REQUIREMENTS CONSUMABLES STORAGE SUPPLY GASES GAS SUPPLY REQUIREMENTS CYLINDER REQUIREMENTS REQUIRED FITTINGS CYLINDER CONNECTIONS BY REGION 35...
The Markforged Metal X System consists of three components: the Metal X printer, the Wash-1 debinding station, and a Markforged furnace (Sinter-1 or Sinter-2). The Metal X printer will arrive at your facility fully assembled and will require limited initial setup. The Wash-1, Sinter-1, and Sinter-2 require additional facilities modification and site preparation by authorized facilities maintenance staff and a certified electrician.
METAL X SYSTEM FACILITIES GUIDE OVERVIEW ELECTRICAL ACCESS The Metal X System installation must provide space for easy access to all electrical disconnecting devices: ● Power switch on rear face of Metal X printer ● Power cable to Wash-1 ● Power switch on furnace (Sinter-1: rear panel, Sinter-2: front panel) ●...
Updates should be performed at the earliest convenient time. NETWORK REQUIREMENTS Markforged devices communicate on the 2.4GHz Wi-Fi band. For assistance with setting up your network, contact your internal IT team. Note that Markforged devices do not support 5GHz Wi-Fi connections. WIRED CONNECTIONS Ethernet connections must utilize cables with a maximum length of 10 meters.
Page 8
DEVICE TIME SYNCHRONIZATION Markforged products require accurate system time for securing HTTPS connections to Eiger and other services; the SSL certificates used to ensure the authenticity of these services have time-limited validity, and the device’s time must fall within this window.
Please locate the appropriate voltage and current combination for your use case based on availability and local regulations. Note: The Metal X can accept a voltage input from 100-120VAC or 200-240VAC. No additional equipment is required. It will automatically detect the input voltage and respond appropriately.
Note: Customer may purchase additional 15-amp rated power cord to utilize alternate receptacle. Power entry module: IEC 60320 Type C20. This list is not exhaustive. It is regularly updated as Metal X System shipments increase. MetalX_Facilities_Guide_2.4 | June 25, 2020...
VENTILATION REQUIREMENTS The Metal X melts thermoplastics and emits thermoplastic odors during printing. Ensure that the Metal X is used in a well-ventilated area. No dedicated exhaust is required for the printer. Additional measures may be required based on local regulations.
MOVING INFORMATION 3m (10ft) Moving the Metal X after installation is not recommended. If you need to move the printer, ensure that all material is unloaded, then remove spools and spool holders. In addition, manually check whether the wiping station is loose and tighten its screws if necessary.
METAL X OPERATIONAL SPACE PLANNING Ensure that the new location of the Metal X meets all facility requirements, including electrical and table requirements; additionally, ensure that the Metal X is placed on the table with at least six inches of table space on all four sides of the printer.
● Ensure that the table can support 1.5x the full load weight of the printer (75 kg / 165 lbs) ● Place the Metal X on a stable surface and take steps to dampen vibration, as excessive vibration may degrade system performance ●...
METAL X SYSTEM FACILITIES GUIDE WASH-1 ELECTRICAL REQUIREMENTS Circuit Type Dedicated branch circuit Wiring System Single phase with safety ground – Input Frequency 60 Hz – Input Voltage 120 VAC Rated Continuous Operating Current 11 A Note: When necessary, the Wash-1 ships with a transformer appropriate to the install region.
METAL X SYSTEM FACILITIES GUIDE WASH-1 INSTALLATION PLANNING Crate Dimensions 0.78m x 0.91m x 1.22m (31in x 36in x 48in) Crate Weight 195 kg (430 lbs) Minimum Door Width to Move in Crate 0.92m (36in) of clearance SHIPPING AND UNLOADING ●...
METAL X SYSTEM FACILITIES GUIDE WASH-1 OPERATIONAL SPACE PLANNING Wash-1Dimensions 0.68m x 0.61m x 1.10m (27in x 24in x 42in) Wash-1Weight 118 kg (260 lbs) Full Load Weight 158 kg (350 lbs) ● The Wash-1 should sit directly on the floor ●...
NOTE ON CONNECTIVITY The Wash-1 is a static debinding unit that does not connect to the Internet. Unlike other Metal X System devices, the Wash-1 does not need to be registered to Eiger or receive firmware updates. LOCAL SOLVENT REGULATIONS Consult local regulations to determine which approved solvents are allowed in your region.
Sinter-1 and Sinter-2. It is expected that you are planning to install one type of device or the other in your facility as part of the Metal X System, not both at once. As such, you are only responsible for the section pertaining to the furnace you plan to install.
If the system is installed with a removable plug, it must be rewired with a hardwired connection per this Facilities Guide; otherwise, Markforged cannot support install or operation. dedicated electrical disconnecting device must be installed for the Sinter-1 as part of the dedicated electrical branch circuit.
SINTER-1 GROUND FAULT PROTECTION IMPORTANTNOTE: The Markforged Sinter-1 complies with IEC, UL, and CSA 22.2 61010-1:2010 and 61010-2-010:2014 standards. Like many industrial furnaces, the Sinter-1 will generate ground current in normal operation. This is expected and acceptable according to the referenced standards.
METAL X SYSTEM FACILITIES GUIDE SINTER-1 CONSUMABLES STORAGE Store all consumables for the Sinter-1 in a cool dry place and leave them in their original packaging. SUPPLY GASES ● A wall-mounted rack with capacity for three or more cylinders should be installed...
METAL X SYSTEM FACILITIES GUIDE SINTER-1 REQUIRED FITTINGS ● Markforged will supply cylinder fittings for certain countries, as detailed below ● For countries where fittings are not supplied, the customer is responsible for supplying appropriate fittings CYLINDER CONNECTIONS BY REGION...
METAL X SYSTEM FACILITIES GUIDE SINTER-1 INSTALLATION PLANNING SHIPPING AND UNLOADING Crate Dimensions 1.32m x 1.10m x 1.50m (54in x 43in x 58in) Crate Weight 225kg (500 lbs) Minimum Door Width to Move in Crate 1.52m (60in) clearance ● Allow for an area of at least 3m x 3m x 1.5m (10ft x 10ft x 5ft) to uncrate the furnace ●...
METAL X SYSTEM FACILITIES GUIDE SINTER-1 ● The centerline of each strap should be 175mm from each end of the furnace ● The straps must be a minimum of 1.8m (6ft) long, though 2.4m (8ft) straps are recommended ● The use of corner protection to avoid abrasion damage is encouraged The Sinter-1 is not intended to be lifted directly from below via forklift as there is insufficient clearance under the unit.
METAL X SYSTEM FACILITIES GUIDE SINTER-1 ACCESS REQUIREMENTS Connectivity Access 150mm (6in) of clearance at rear of furnace (Wi-Fi, Ethernet) 150mm (6in) at rear of furnace USB Access 150mm (6in) on left side of furnace Overhead Clearance No flammable materials within 920mm (36in)
METAL X SYSTEM FACILITIES GUIDE SINTER-1 EMERGENCY PLANNING Emergencies with the Sinter-1 are unlikely, but it is recommended that facilities consider the following as a precaution. FIRE EXTINGUISHERS It is recommended that facilities have a multipurpose fire extinguisher near Sinter-1.
The system cannot be installed with a removable plug. A unit installed with a removable plug must be rewired with a hardwired connection per this Facilities Guide; otherwise, Markforged cannot support the install or operation. The hardwired connection cable must be long enough to allow some movement of the Sinter-2 unit during operation and maintenance.
Sinter-2 and be readily accessible at all times: ● A safety switch compliant with local codes and regulations ● A circuit breaker compliant with local codes and regulations Markforged recommends a heavy-duty disconnect switch as the dedicated Sinter-2 electrical disconnect. MetalX_Facilities_Guide_2.4 | June 25, 2020...
METAL X SYSTEM FACILITIES GUIDE SINTER-2 ENVIRONMENTAL REQUIREMENTS CLIMATE CONTROL Intended Use Indoor use, floor use only Room Temperature (Operation) 18.8°C – 29.4°C (66°F – 85°F) Humidity Requirement (Operation) 0% – 95%, non-condensing Room Temperature (Idle) 18.8°C – 29.4°C (66°F – 85°F) 0% –...
METAL X SYSTEM FACILITIES GUIDE SINTER-2 CONSUMABLES STORAGE Store all consumables for the Sinter-2 in a cool dry place and in their original packaging. SUPPLY GASES ● A wall-mounted rack with capacity for three or more cylinders should be installed...
METAL X SYSTEM FACILITIES GUIDE SINTER-2 REQUIRED FITTINGS ● Markforged will supply cylinder fittings for certain countries, as detailed below ● For countries where fittings are not supplied, the customer is responsible for supplying appropriate fittings CYLINDER CONNECTIONS BY REGION...
METAL X SYSTEM FACILITIES GUIDE SINTER-2 INSTALLATION PLANNING SHIPPING AND UNLOADING The Sinter-2 arrives in crates: one containing the furnace and gas supply piping, and one accessory crate containing all other parts and equipment (including the furnace tube). The furnace must be moved into its final operating location before the Installer arrives onsite.
METAL X SYSTEM FACILITIES GUIDE SINTER-2 MAINTENANCE SPACE PLANNING Connectivity Access 150mm (6in) of clearance at rear of furnace (Ethernet) 150mm (6in) at rear of furnace USB Access 150mm (6in) on right side of furnace Overhead Clearance No flammable materials within 920mm (36in)
METAL X SYSTEM FACILITIES GUIDE SINTER-2 NETWORK CONNECTIVITY The Sinter-2 can only connect to a local network over Ethernet. It therefore requires an open Ethernet jack and cable for Internet access. Wi-Fi is not available on Sinter-2. EMERGENCY PLANNING Emergencies with the Sinter-2 are unlikely, but it is recommended that facilities consider the following as a precaution.