• Do not reuse the test strips, test strips are single use only.
• You may use this system at altitude up to 10,000 feet.
• Dispose of used test strips and lancets according to instructions from your
healthcare provider.
• The ProView Blood Glucose Monitoring System is for testing outside of the body
(in-vitro diagnostic use).
• Inaccurate results may occur for individuals or patient in shock.
• Inaccurate results may occur for individuals experiencing a hyperglycemic
hyperosmolar state with or without ketosis. Critically ill patients should not be
tested with this blood glucose meter.
• Interferences: Acetaminophen, salicylates, uric acid, ascorbic acid (vitamin C) and
other interfering substances in normal blood or normal therapeutic concentrations
do not significantly affect results, however abnormally high concentrations in blood
may cause inaccurate results.
• Testing under out of the specification range may cause inaccurate results.
• Patients undergoing oxygen therapy may yield false low results.
• Lipemic samples (triglycerides) in excess of 1,500 mg/dL may produce elevated
results.