1-4.
Optional Accessories
The optional accessory of this unit is as follows:
• Foot Switch FS-24
Note
FS-24 is drip-proof protection class IPx3. Therefore, FS-24 can not be used in the place that is exposed to a liquid like
an operating room. To safety, in case the product is used in the place that is exposed to a liquid like an operating room,
use the product that is compatible with IPx6 or higher.
REMOTE contact point switch connector
Connector specifications (stereo miniature jack)
Pin No.
Signal Name
1
GND
2
TRIGER SIGNAL (TTL)
3
BUSY STATUS (TTL)
HVO-500MD/HVO-550MD
3
2
1
Function
–
When the condition is changed more than 100 msec, the action is started
according to the setting of the REMOTE contact point switch.
According to the setting of the REMOTE contact point switch connector,
it becomes high level.