Download Print this page

ONOSOKKI FT-2500 Manual page 3

Advertisement

Protection against the electrical noise
Protection against the electrical noise
Parts list
Parts name
Manufacturer
Line filter
Schaffner Holding AG
Surge killer
Surge killer
Phoenix Contact GmbH & Co.KG
Surge killer
(Germany)
Base for surge killer
Outlet
AC
Line
AC
IN
filter
IN
Surge killer
Measurement panel
Digital IO
Connector for remote input, comparator output and pulse output
Pin NO.
Functions
1
Comparator UPPER
2
3
Comparator LOWER
4
5
Comparator ROTATION
6
7
Comparator OK
8
9 to 11
Not connected
12
Pulse output SIG
13
Pulse output COM
14
Remote input SIG
15
Remote input COM
(Note) Pulse output signal (the above No. 12/13) is equivalent
to that of the displaying of the frequency.
Remote input
Remote input
FT-2500 Input circuit
+5V
+5V
Remote input terminal
MODE
Open
Close
NORMAL
Measure
Hold
REVERSE
Hold
Measure
Pulse output
Pulse output
FT-2500 Output circuit
+5V
5kΩ
500Ω
The frequency of measured power spectrum is output after being
converted to the pulse signal. Therefore, the displayed value may
differ from the frequency of pulse output when the rotational
speed is selected as a display.
Comparator output
Comparator output
FT-2500 Output circuit
Photo-MOS relay enables the FT-2500 to connect directly to PLC* etc.
*PLC : Programmable Logic Controller
A d v a n c e d Ta c h o m e t e r
Model name
FN9222R-3-06
F-MS 12ST
VAL-MS 230ST
VAL-MS 230ST
VAL-MS-BE
Power cable
FT-2500
Sensor
Signal cable
Protective grounding
Protective grounding
DIGITAL I/O
SIG-1
Load resistance
100kΩ or more
FT-2500
When installing the FT-2500, the following precautions should be taken care of.
Some installation conditions may give adverse influence against the noise tolerance.
Separate the power supply cable of the FT-2500 from the power line which is connected to high-power
load.
Be sure to use the power supply cable which is provided as standard accessory.
Do not arrange the wiring of the FT-2500 cables in parallel or together with the power line.
Do not extend the sensor signal cable longer than necessary.
Use cables of 5m or less in length for DIGITAL-I/O and V-OUT.
Use a shielded cable as the signal cable. In addition, be sure to ground the shielding wire.
Keep the FT-2500 as far away as possible from devices, which are generating the strong high-frequency
signal or surge.
Keep the FT-2500 and its cables away from devices, which are generating the strong electric and
magnetic fields.
Be sure to connect the FT-2500 to protective ground.
When installing the FT-2500 inside a control or measurement panel, ground the instrument shielding wire
to the panel and also ground the control or measurement panel.
If it is subject to influences by electrical strong noise or surge, use a surge killer and noise filter inside the
control or measurement panel as shown in the figure left.
*
It is requested to wire the signal cable as short as possible. Keep the minus side of the surge killer within
50 cm. Ground both ends of the shielding wires of all input/output signal cables to the ground terminal of
the panel.
RS-232C communication
RS-232C communication can be used
with the AX-5022 signal cable.
RS-232C
V-OUT
SIG-2
100-200 V
50/60 Hz
22-32 VA
LAN (Ethernet option) FT-0251
The FT-2500 can be connected to LAN by using Ethernet.
(Note) LAN and RS-232C communication cannot be used at the same time.
V-OUT connector
Connector for analog voltage output.
It can output the analog voltage signal for input signal
monitoring depending on the setting.
REVO
Voltage output in proportion to the rotational speed
SIG
Signal output for the sensor's signal monitoring
Pin No.
A
SIG
B
COM
C
Not connected
(Note) The signal after envelope-processed and sensitivity adjusted is output from SIG.
Connector for sensor input
Select appropriate input connector from SIG-1 or SIG-2
depending on the sensor.
SIG1 : FT-0501
SIG2 : FT-0801, IP-292, IP-296, IP-3000A, IP-3100, OM-1200,
OM-1500, VP-202, VP-1220, NP-3000 series, MI series
POWER
Network I/F
100BASE-TX/10BASE-T (automatic switching)
Protocol
TCP/IP
Connector
RJ-45
Function
Rear
3

Advertisement

loading