Handheld Advanced Tachometer
FT-7100
Instruction Manual (Function Reference)
Thank you for your selection of the FT-7100 Handheld
Advanced Tachometer.
To ensure the performance of the FT-7100, please read
this manual thoroughly.
Copyright © ONO SOKKI CO.,LTD.
2007 All rights reserved.
■ Omission of Issuance of Certificate
This product has been tested under strict inspections for
correct operation before shipment. Please note that the
issuance of certificate is omitted.
■ Warranty
1. This product is covered by a warranty for a period of one
year from the date of delivery.
2. This warranty covers free-of-charge repair during the
warranty period for defects occurred while the product is
used under correct operating conditions according to de-
scriptions in this manual and notices on the unit label.
3. For free-of-charge repair during the warranty period, con-
tact your dealer or your nearest Ono Sokki sales office
nearby.
4. Even during the warranty period, the following failures
will be handled on a fee basis.
(a) Failures or damages occurring through misuse,
misoperation, repairing without ONO SOKKI'S ap-
proval.
(b) Failures or damages occurring through mishandling
(dropping) during transportation after purchase.
(c) Failures or damages occurring by an Act of God (fires,
earthquakes, flooding, and lightening), environmen-
tal disruption, or abnormal voltage.
(d) Replenishment of expendable supplies, spare parts,
and accessories.
This guarantee covers only the performance of the product itself only.
All inconvenience by the trouble of this product is not included.
*Outer appearance and specifications are subject to change without prior notice.
HOME PAGE: http://www.onosokki.co.jp/English/english.htm
WORLDWIDE
Ono Sokki Co., Ltd.
1-16-1 Hakusan, Midori-ku,
Yokohama 226-8507, Japan
Phone : 045-935-3976
Fax : 045-930-1906
E-mail : overseas@onosokki.co.jp
B00002055 / IM07101502(2)07X(MS)XXX
Functions and Operations
1.Power Switch
① When you slide the power switch upward, the power of the main
unit turns ON.
② When you turn ON the power, the software version is displayed
in the MAIN display and product code "FT7" of the main unit
in the SUB display. Then, the measurement mode is entered.
③ Each parameter condition used in previous measurement is
backed up.
Measurement Mode
Power switch
Ends the measurement mode and then turns OFF the power.
At the time of rotational acceleration/deceleration
RECALL &
↑ switch
measurement (when Algorithm E of the ACT mode is
selected), this switch is used to select a rotational
speed at the start of measurement from sampled
candidate values.
Selects the parameter setup mode.
MENU switch
In the measurement mode, at the time of rotational
SAMPLE &
→ switch
acceleration/deceleration measurement (when the
"ACT" mode is selected), this switch is used to sample
rotational speed data before starting measurement and
calculate candidate values of the rotational speed at the
start of measurement.
At the time of rotational acceleration/deceleration
MODE &
measurement (when Algorithm E of the ACT mode is
NEXT switch
selected), this switch is used to determine a value selected
by the RECALL & ↑ switch as a rotational speed at the start
of measurement, and then starts measurement.
Note: When Algorithm D is selected, the function to select a candidate value of the rotational speed is not provided.
3.Setup Mode
When you press the MENU switch in the measurement mode, the parameter setup mode is selected.
Then, set parameters using the RECALL & ↑ and SAMPLE & → switches. Apply parameters and select items using the MODE &
NEXT switch.
Operation flow of the parameter setup mode is shown below.
Measurement mode
MENU
Setup mode
Sensor
selector switch
When OTHERS
When IP is selected
is selected
RECALL
SAMPLE
&
Set each parameter
SENSOR-SEL ( SNS )
SENSOR-SEL ( SNS )
[ IP-292 : IP-296 : IP-30 ]
[ OTHER : NP ]
MODE
Apply &
select item
ENVELOPE-SEL ( ENV )
NEXT
[ ON : OFF ]
INPUT-LEVEL-SEL ( LVL )
[ 5V : 0.5V : 0.05V ]
MEASUREMENT-MODE-SEL ( MOD )
[ CNS : ACT ]
When CNS is selected
When ACT is selected
ALGORITHM-SEL ( ALG )
ALGORITHM-SEL ( ALG )
[ TYPE-A : TYPE-B ]
[ TYPE-D : TYPE-E ]
AVERAGE-SEL ( AVE )
[ OFF : 2 : 4 : 8 : 16 ]
FREQUENCY-RANGE-SEL ( RNG )
[ 2000 : 500 : 250 ]
④ When you perform measurement for the first time, set each
parameter first.
2.Function of Each Switch
When you turn ON the power, each switch has a different func-
tion between the measurement mode and the parameter setup
mode.
The function of each switch in each mode is shown below.
Parameter Setup Mode
Cancels the current setting and then turns OFF the power.
Changes the selection of the current setting. During
numerical parameter setup, increments the numerical
value of the relevant digit. When the value is 9, returns
to 0.
Applies the current setting and then changes to the
measurement mode.
During numerical parameter setup, moves the setting
cursor to the right. When the setting cursor is at the least
significant digit, returns to the most significant digit.
Applies the current setting and then moves to the next
setting.
Operate the sensor selector switch before setting parameters
in the setup mode.
MENU
Terminate the constant
setup mode
BACK-LIGHT ( LGT )
[ ON : OFF ]
CAL-SEL ( CAL )
[ 0 V : 1 V ]
ANALOG-FS-SET ( FS )
[ 1∼99999 ]
When REV is selected
When SIG is selected
MONITOR-SEL ( MON )
[ REVO: SIG ]
PULSE-SET ( PLS )
[ 0.5∼200.0 ]
FILTER-UPPER-SET ( UPP )
[ 1∼99999 ]
FILTER-LOWER-SET ( LOW )
[ 0∼99998 ]
When ON is selected
When OFF is selected
FILTER-SEL ( FIL )
[ ON : OFF ]
Selecting a sensor to be connected (Sensor → SNS)
Select a sensor to be used.
First, select the type of a sensor ("IP" or "OTHERS") to be
used using the sensor selector switch on the right-hand side
of the main unit. Then, select a sensor to be used in the setup
mode.
IP
IP
When an IP Series sensor is used
I
OTHERS
OTHERS
When a sensor other than IP Series is used
IP292 Select when IP-292 is used
IP
IP296
Select when IP-296 is used
IP-30 Select when IP-3000A or IP-3100 is used
nP
Select when an NP Series sensor is used
OTHERS
OtHEr
Select when a sensor other than above is used
* Set to "OtHEr" at the time of shipment.
Note: Connecting a sensor of voltage signal output type with
IP-292 selected may cause damage to the sensor. Be
sure to select "OTHERS" before connecting the sen-
sor.
Selecting envelope processing (Envelope → ENU)
Select the presence or absence of envelope processing.
This item can be set only when "nP" or "OtHEr" is selected in "①
Selecting a sensor to be connected."
OFF
Without envelope processing
On
With envelope processing
* Set to "OFF" at the time of shipment.
③ Selecting the input voltage level (LEVEL
LUL)
Set the input voltage level.
Use together the sensor amplifier sensitivity adjustment
volume on the right-hand side of the main unit to adjust
the input voltage level so that the indicator stably blinks.
Max. input amplitude ±5 V
5u
Max. input amplitude ±0.5 V
0.5u
Max. input amplitude ±0.05 V
0.05u
* Set to "5u" at the time of shipment.
Setting the drive mode (Mode → Mod)
Set the drive mode.
CnS
Used when fluctuations of the rotational
speed of the object under measurement
are small (when measuring rated
rotational speed, etc.)
Used when measuring the rotational
ACt
speed of an object under measurement,
which is accelerating or decelerating.
However, measurement may not be
performed correctly if the speed
changes rapidly.
* Set to "CnS" at the time of shipment.
Setting a measurement algorithm (Algorithm → ALG)
Set a measurement algorithm.
When the drive mode is set to "CnS", only "tYP-A" and "tYPb"
can be selected.
When the drive mode is set to "ACt", only "tYP-d" and "tYP-E"
can be selected.
For explanations of each algorithm, refer to "Descriptions of
Each Algorithm" in <Various Measurement Operations>.
tYP-A
Select Algorithm A
tYP-b
Select Algorithm B
tYP-d
Select Algorithm D
tYP-E
Select Algorithm E
* Set to "tYP-A" at the time of shipment.
Setting the number of moving averagings (Average → AUE)
Set the number of moving averagings.
OFF Does not perform the moving average.
2
Performs moving averaging with the latest two data items.
4
Performs moving averaging with the latest four data items.
8
Performs moving averaging with the latest eight data items.
16
Performs moving averaging with the latest 16 data items.
* Set to "OFF" at the time of shipment.
Setting the frequency range (Range → RNG)
Set the frequency range.
2000
Set the input frequency range to 30 to 2000 Hz.
500
Set the input frequency range to 7.5 to 500 Hz.
250
Set the input frequency range to 3.75 to 250 Hz.
* Set to "250" at the time of shipment.
Selecting the filter function (Filter → FLT)
Set the filter function to On or OFF .
When set to On, setups
and
become possible.
OFF
Without filter function
On
With filter function
* Set to "OFF" at the time of shipment.
Setting the lower limit of filter range (Lower → LOW)
Set the lower-limit rotational speed (r/min) of the filter range.
This item can be set only when the filter function is set to
On.
Setup range: 0 to 99998
* Set to "00000" at the time of shipment.
Note: When Algorithm B is selected, the filter function is
disabled.
Setting the upper limit of filter range (Upper → UPP)
Set the upper-limit rotational speed (r/min) of the filter range.
This item can be set only when the filter function is set to
On.
Setup range: 1 to 99999
* Set to "99999" at the time of shipment.
Note: When Algorithm B is selected, the filter function is
disabled.
Setting the number of pulses (Pulse → PLS)
Set the number of pulses per rotation (P/R) according to the
object under measurement.
Setup range: 0.5 to 200.0
Step: 0.5
* Set to "001.0" at the time of shipment.
* To be continued to backside.