Rear Panel Pin Outs - Cardionics SimulScope Operator's Manual

Table of Contents

Advertisement

ACCESSORY OUTPUTS:
SOUNDS:
ECG (option):
POWER:
BATTERY:
BATTERY LIFE:
POWER/LOW BATTERY: Green indicator for power on/off. Turns orange
INDICATOR:
AC POWER SUPPLY/
BATTERY CHARGER:
SAFETY:
___________________________________________________________

Rear Panel Pin Outs

ACCESSORY
2
1
Specifications (Cont.)
3VppK nominal, 9VppK, max. 1K ohm output impedance.
FM modulated, 4.5 KHz center frequency, 30 Vppk output
for tape recorder.
Nickel Cadmium battery pack w/external access for
replacement.
2.5-3 hours continuous operating time on fully recharged battery.
to indicate low battery condition with approximately 30
minutes operating time remaining.
Wall mount 120VAC power supply allows unit to operate
while charging battery. Fully depleted battery
requires 12-16 hours for recharge.
Meets ANSI/AAMI standard, safe current limits
for electromedical apparatus, non-isolated patient connection.
3
Function
-----
Sounds out
Ground
4
----------
Vext
6
ECG
Pin
Function
1
Steth In
2
Ground 2
3
+12V
4
----------
5
----------
6
----------
27
STETH IN
5
3
4
1