JUMO 902454/10 Operating Manual

JUMO 902454/10 Operating Manual

Temperature probes for combined cold/heat meters

Advertisement

Quick Links

Temperature Probes
for Combined Cold/Heat Meters
Basic types 902454/10 and 902454/11
Basic type 902464/10
Operating Manual
90245400T90Z001K000
V1.00/EN/00731587/2020-01-27

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 902454/10 and is the answer not in the manual?

Questions and answers

Summary of Contents for JUMO 902454/10

  • Page 1 Temperature Probes for Combined Cold/Heat Meters Basic types 902454/10 and 902454/11 Basic type 902464/10 Operating Manual 90245400T90Z001K000 V1.00/EN/00731587/2020-01-27...
  • Page 3: Table Of Contents

    Installation notes ............12 Temperature probes for direct mounting (902454/10 and 902454/11) ....13 Temperature probes in thermowells (902464/10) .
  • Page 4 Contents...
  • Page 5: Safety Information

    Safety information 1 Safety information General This manual contains information that must be observed in the interest of your own safety and to avoid material damage. This information is supported by symbols which are used in this manual as indicated. Please read this manual before starting up the device.
  • Page 6 1 Safety information...
  • Page 7: General Information

    0 °C). The nominal values are normally 100 Ω (Pt100), 500 Ω (Pt500), and 1000 Ω (Pt1000). The RTD temperature probes in the type series 902454/10 and 902454/11 for direct mounting, and 902464/10 for installation in thermowells have national type approval for cold meters. When it comes to use on combined cold/heat meters, these temperature probes also have an EC type examination in ac- cordance with EC Directive 2014/32/EU (MID) including Appendices I and MI-004.
  • Page 8 2 General information...
  • Page 9: Technical Data

    Technical data 3 Technical data Temperature range 902454/10 0 to 120 °C 902454/11 0 to 120 °C 902464/10 0 to 120 °C The maximum operating temperature of the thermowells must be observed. Protection type IP65 (as delivered condition) Temperature difference...
  • Page 10 3 Technical data...
  • Page 11: Installation

    Installation 4 Installation If the pair of temperature probes is connected to a computer unit, make sure that the probe's nominal value matches that of the processing computer unit. Furthermore, make sure that the installation location is deep enough to prevent damage to the tip of the probe or thermowell when screwing in.
  • Page 12: Installation Notes

    4 Installation Installation notes The cold barrier serves the thermal decoupling of the terminal head. This prevents condensation forming in the terminal head. Condensation forms on surfaces with a temperature below the dew point tempera- ture of the surrounding ambient air. In a normal room environment of 20 °C and 50 % relative humidity, the dew point temperature is approx.
  • Page 13: Temperature Probes For Direct Mounting (902454/10 And 902454/11)

    4 Installation Temperature probes for direct mounting (902454/10 and 902454/11) Screw in only at hexagonal head Direct mounting Pipe Ø 50 mm 60.3 mm 85 mm -39 mm 40 mm 65 mm 76.1 mm 120 mm -54 mm 60 mm 80 mm 88.9 mm...
  • Page 14: Temperature Probes In Thermowells (902464/10)

    4 Installation Temperature probes in thermowells (902464/10) Spanner flat for mounting Installation in thermowell Pipe Ø 150 mm 168.3 mm 140 mm -73 mm 40 mm 120 mm 200 mm 219.1 mm 220 mm -131 mm 70 mm 210 mm 250 mm 273.0 mm 220 mm...
  • Page 15 This version complies with the DIN EN 1434-2 standard and is preferred. The thermowell is marked by an appropriate DIN EN 1434 standard reference and the insertion length, as well as pressure stage, in- ternal diameter, maximum operating temperature, and JUMO company logo. Version 2 Thermowell internal diameter: (6.2 +0.00/-0.05) mm...
  • Page 16 4 Installation...
  • Page 17: Maintenance

    Maintenance 5 Maintenance In order to maintain measurement stability, a metrological inspection must be carried out when the cali- bration period/the metrological identification marking has elapsed to check that the maximum permissi- ble error (MPE) is observed. At the same time, to maintain leak-tightness, at least every 5 years the seal on the cable entry and the terminal head lid must be replaced.
  • Page 18 5 Maintenance...
  • Page 19: Declaration Of Conformity

    Declaration of conformity 6 Declaration of conformity...
  • Page 20 6 Declaration of conformity...
  • Page 21: China Rohs

    China RoHS 7 China RoHS 3URGXFW JURXS   &KLQD ((3 +D]DUGRXV 6XEVWDQFHV ,QIRUPDWLRQ &RPSRQHQW 1DPH +J &G &U 9, 3%'( +RXVLQJ *HKlXVH 3URFHVV FRQQHFWLRQ 3UR]HVVDQVFKOXVV 1XWV 0XWWHU 6FUHZ 6FKUDXEH ^: d 7KLV WDEOH LV SUHSDUHG LQ DFFRUGDQFH ZLWK WKH SURYLVLRQV 6-7  *%7  ,QGLFDWH WKH KD]DUGRXV VXEVWDQFHV LQ DOO KRPRJHQHRXV PDWHULDOV¶...
  • Page 22 7 China RoHS 3URGXFW JURXS  &KLQD ((3 +D]DUGRXV 6XEVWDQFHV ,QIRUPDWLRQ &RPSRQHQW 1DPH +J &G &U 9, 3%'( +RXVLQJ *HKlXVH 3URFHVV FRQQHFWLRQ 3UR]HVVDQVFKOXVV 1XWV 0XWWHU 6FUHZ 6FKUDXEH ^: d 7KLV WDEOH LV SUHSDUHG LQ DFFRUGDQFH ZLWK WKH SURYLVLRQV 6-7  *%7  ,QGLFDWH WKH KD]DUGRXV VXEVWDQFHV LQ DOO KRPRJHQHRXV PDWHULDOV¶...
  • Page 24 JUMO GmbH & Co. KG JUMO Instrument Co. Ltd. JUMO Process Control, Inc. Street address: JUMO House 6733 Myers Road Moritz-Juchheim-Straße 1 Temple Bank, Riverway East Syracuse, NY 13057, USA 36039 Fulda, Germany Harlow, Essex, CM20 2DY, UK Delivery address:...

This manual is also suitable for:

902454/11902464/10

Table of Contents