PERFORMANCE CHARACTERISTICS
Within run precision
(100 venous blood tests per glucose level)
Concentration
30~50 mg/dL
51~110 mg/dL
111~150 mg/dL
151~250 mg/dL
251~400 mg/dL
Total precision
(100 control solution tests per glucose level)
Concentration
30~50 mg/dL
90~130 mg/dL
280~420 mg/dL
38
2in1.® BASIC
Standard deviation
(mg/dL)
1.8
3.1
5.8
6.1
5.8
2in1.® BASIC
Standard
deviation (mg/dL)
2.1
2.8
6.0
System
Coeffi cient of variation
(%)
3.7
2.9
4.4
2.6
1.6
System
Coeffi cient of
variation (%)
4.3
2.7
1.8