ST2200 Soldering Iron Tester User Manual
Move equipment to another AC outlet and repeat the above tests. If the ground
has been verified and the ST2200 sensor has been replaced but the over range
indication of "1---" still exists, return the ST2200 to the factory for evaluation and
repair.
8.5.4 Troubleshooting - Wrist Strap Test Problems
If a wrist strap under test indicates a failure, following are possible causes.
1. Check that the wrist strap cable connector is connected properly to the
ST2200. See Figure 1, Item 9 for wrist strap connector location. Do not rely on
common ground of AC line for wrist strap testing.
2. Make sure the operator is not touching any other conductive material. Ensure
that the operator's free hand is not in contact with any other object.
3. Make sure wrist strap is in contact with the skin. Do not wear the wrist strap
over clothing.
4. Wrist strap may be defective, replace wrist strap.