3.7. OPENING THE ENCLOSURE

The instructions detailed within the product certificate and section 7 CERTIFICATION APPENDIX
of this manual must be followed at all times.
1.
WARNING – DO NOT OPEN WHEN ENERGIZED.
2.
Using a 3mm AF allen key loosen the M4 locking screw located
at the 12 o'clock position on the rear endcap 6x full turns.
3.
Carefully unscrew the rear endcap while being prepared to take
the weight when it disengages. There are 4x unthreaded holes
around the perimeter of the endcap which may be used should
leverage be required.

3.8. CLOSING THE ENCLOSURE

1.
Gently hand tighten the rear endcap. Do not use tools or over
tighten otherwise the endcap will be difficult to remove. The M4
locking screw should be at the 12 o'clock position when the
endcap is fully engaged, there should still be a small gap
(<0.5mm) between the endcap and housing.
2.
Using a 3mm AF allen key gently tighten the M4 locking screw
located at the 12 o'clock position on the rear endcap. Do not
over tighten, very little torque is required to lock the endcap in
place.