Therm-App MD User Manual
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------
WARNING
Be careful when you clean the infrared lens. The lens has a delicate anti-reflective coating. Do not clean the lens too vigorously; this can
damage the anti-reflective coating.
WARNING
The Product allows for field replacement of the lens. Exchanging the lens should be done with great caution, as after the removal of the
lens the IR detector is exposed until the new lens is installed. While exposed, avoid any contact with the detector. Touching, cleaning or
blowing on the detector could cause permanent damage. Replacement of the lens should be done in a dry area, using clean hands with
maximum care.
WARNING
Do not use pressurized air from pneumatic air circuits in a workshop to remove dust from the detector. This air contains oil mist intended
to lubricate pneumatic tools and is set a at a level of pressure that is too high. This can damage the detector and lens.
WARNING
The device is designed to use 5V DC power supply only, fed through its micro USB connector. Using any other type of power source is
dangerous and may lead to damage.
WARNING
The Product is provided with appropriate USB-OTG cables. Use only the supplied USB OTG cables. Using any other types of cables
might not work at all, may fail, and even damage the equipment.
------------------------------------------------------------------------------------------------------------------------------------------------------------