The sample has been measured using the B (blue) filter. The Phosphorescence (emitted light
intensity) measured after the Cool down time is 21.2% of the Fluorescence, measured after the
Exposure Time before switching the light source off. The charging time is estimated to be at least
63.9ms to reach saturation. The discharging time is estimated to be 19ms. After 0.4ms the
Phosphorescence decays below 50% of the Fluorescence. After 6.6ms the Phosphorescence
decays below 20%.

MSpez function

In this mode you can measure the delta E between two samples
Measure the Target several times. The Average is taken as reference.
Key A: Switch to sample measurement
Key B: Positioning LED on
Key D: Exit
Measure the sample several times. The Average is calculated.
Key A: New reference