INTENDED USE
The Fingertip Pulse Oximeter is indicated for
use in measuring oxygen saturation of arterial
hemoglobin (SpO
) and pulse rate. It is intended
2
for patients during no-motion condition. The
patients are limited to adults with weight above
88 lb. This device is indicated for non-invasive
spot checking or monitoring.
METER OVERVIEW
Principle of Measurement
The Fingertip Pulse Oximeter determines
functional oxygen saturation of arterial
hemoglobin (SpO
) by measuring the absorption
2
of red and infrared light passing through perfused
tissue. Changes in absorption caused by the
pulsation of blood in the vascular bed are used to
determine oxygen saturation and pulse rate.
6
Meter Appearance and Key
Function
1
2
3
4
5
6
7
8
Bluetooth Indicator
1
Blue light appears when bluetooth is turned on.
SpO
%
2
2
The measurement result of oxygen saturation
in percentage.
3
Battery Indicator
Pulse Amplitude
4
The strength of the signal is detected by the
oximeter.
Backlight (White or Red)
5
Backlight is white while in measuring mode.
Backlight is blinking red while the oxygen
saturation value is below 85%. (high priority
visual alarm)
6
Pulse Rate
The measurement result of pulse rate in beats
per minute.
7
On/Off Button
It is used to turn on or turn off the oximeter by
pressing On/Off button
8
Battery Compartment