Maintenance of water trap/filter unit
AFTER EACH ANALYSIS, CHECK FOR ANY PRESENCE OF WATER IN THE CONDENSATE COLLECTION BOWL
AND ELIMINATE IT, IF ANY. PUT THE PROBE BACK IN THE CASE ONLY AFTER YOU HAVE ELIMINATED
CONDENSATE FROM THE TUBE AND THE EXPANSION TANK (SEE CHAPTER 'MAINTENANCE').
REPLACE THE FINE DUST FILTER IF IT IS VISIBLY DIRTY OR WET (SEE CHAPTER 'MAINTENANCE'). DO NOT PERFORM
ANY MEASUREMENT WHEN THE FILTER IS REMOVED OR DIRTY IN ORDER TO AVOID ANY RISK OF IRREVERSIBLE
DAMAGES ON SENSORS.
To remove the water trap, simply turn the cover and remove the filter housing; then remove the inner container and replace the filter
(see figure).
Clean only with water and dry all components of the filter unit thoroughly before reassembling.
Replacing the particulate filter
If the particulate filter has turned black, especially on the outer surface (see example), it needs to be replaced immediately. This is to
avoid any obstacle to the gas flow.
Blackened particle filter