CAUTION: Never actuate the shuttle valve while the hand pump is under pressure or vacuum. Actuate the shuttle valve only when the pump is
vented.
Apply Pressure or Vacuum
1
Verify that the vent valve is open.
2
For positive pressure, turn the fine adjustment knob to the full, counter-clockwise position. For vacuum, turn the fine adjustment knob to the full, clockwise