System Configuration - Keithley S46 Series Installation Manual

Microwave switch system
Hide thumbs Also See for S46 Series:

Advertisement

Figure 4
Relay connector locations
System configuration
After installing the new relays, you must program the S46 for the new relay configuration with the following
command:
:ROUT:CONF:CPOL <clist>
<clist> is defined as follows:
clist = (@0 | 3-6, 0 | 3-6, 0 | 3-6, 0 | 3-6, 0 | 1, ... 0 | 1)
Relay:
A
B
NOTE All switch locations must be included in the <clist> whether or not all
See Section 3 of the S46 Instruction Manual (document number S46-901-01) for detailed programming
information.
C
D
1
... 8
locations are populated. The numeric value (1 or 3-6) indicates the
number of throws, while a value of 0 indicates that a location is not
populated.
For example, if RELAY A and RELAY B are SP6T, RELAY 1 and RELAY
2 are SPDTs, and all other locations are not populated, the command
syntax would be:
:ROUT:CONF:CPOL (@6,6,0,0,1,1,0,0,0,0,0,0)