Display; Function; Heating - Fagron FagronLab PRO-DMS User Manual

Table of Contents

Advertisement

FagronLab™ PRO-DMS

Display

Remote control
Figure 4. Device display
Characters
Temp and °C
Stir
Hot
Probe
Remote
Err
Rated value/Real value
Note:
If both heating and stirring functions have been started at the same time, heating function always
has higher priority. If in this case speed is changed via the stirring knob, it displays stirring speed and
reverses to temperature in the duration of 5 seconds.

9.Function

Heating

The device is controlled by digital temperature control technology, with two separate safe circuits. The hotplate is kept
at a constant temperature by a digital control circuit. The hotplate temperature can also be monitored from a separate,
adjustable safe circuit. The two temperature sensors (PT1000) internal for temperature control are built into the hotplate.
The single external PT1000 can monitor the temperature of the sample.
Real value
Remote
Probe
Hot
External probe
Hotplate warning
Description
Display temperature when the heating function is switched ON
Display stirring state when the stirring function is switched ON
Display hot warning if the heating plate temperature is above 50°C after switch-
ing OFF the heating function
Display when using external probe
Display in case of remote control
Display in case of an error
Display the value of heating or stirring when function is ON
Temperature
Stir
Temp
Stir
C
Err
Set
o
A
B
C
Mode
FagronLab™ PRO DMS – User Manual | 8
Settings
Error
Rated value

Advertisement