CDP Blow & Go 4500 Manual

Coin-operated alcohol breath tester

Advertisement

Quick Links

Blow & Go
CDP 4500
Coin-operated Alcohol Breath Tester
1

Advertisement

loading

Summary of Contents for CDP Blow & Go 4500

  • Page 1 Blow & Go CDP 4500 Coin-operated Alcohol Breath Tester...
  • Page 2 1. Description The Blow & Go coin-operated Alcohol Breath Tester is designed to measure your alcohol concentration level. With an advanced fuel-cell sensor, this device offer high sensitivity and is extremely accurate. Its easy testing process and complete functions make it versatile. When your alcohol blood content exceeds the preset alarm level, it will send an alarm warning and show the result directly and clearly on the LCD digital display.
  • Page 3 3. Installation Guide Screw installation orifice On/Off switch Straw exit Power cable Photo 1 First find a suitable place to install the unit. Then use bolts to fasten the device on the wall according to Photo 1. Note the distance and measurements. 4.
  • Page 4 5. Setting the coin quantity to carry out a test. ● Note: Please cut off the power supply before operation. The number of similar coins required is set by dial switch S1 located inside the unit. This means that the switch on the top is 1 and the switch on the bottom is 0. Setting range from 0-15 units.
  • Page 5: Buttons And Functions

    6. Adjusting the electronic coin selector: See video at this link: http://youtu.be/Sn1ieVK0UL4 How to configure the coin model Turn on the unit, open the front part of the case, you will see LED (7) flashing. Press the setup button (5) during two seconds and then release it. The LED (7) indicator on the top left will change from red to green.
  • Page 6 7. Test procedure 7.1. Mount the device on the wall securely; connect the power cord, then press the switch button on the side. 7.2. Press the On/Off switch on the side of tester. When the tester is on, it will begin to initialize.
  • Page 7 9. Calibration This high-tech sensor-module design does not require complex calibrations. You only need to change the module when the tester flashes the number 0000. Remember that each sensor module can be used 1000 times. Turn off the tester. Open its front panel. (See Picture 8). You will see a yellow ...
  • Page 8: Troubleshooting

    11. Troubleshooting Fault Possible reason Solution No display Unstable power cord Plug in again connection on LCD Circuit fault Contact distributor No response to Warm-up has not concluded Wait until the warm-up gas detection concludes Circuit fault Contact distributor 12. Contact data C.D.Products S.A.