5.8.2 DEMINERALISED WATER QUALITY CONTROL
The B-One autoclave is supplied with an integrated system for input water conductivity
verification.
Each time demineralised water is loaded, the autoclave verifies if loaded water is suitable
for use.
The user will be notified if water is not suitable: the corresponding icon on the display
signals that the loaded water does not have suitable features for sterilization, and the lit
ring on the knob will turn violet.
This same icon is also located over the front outlet which allows loading tank
draining.
When this icon appears, the user should empty the tank and top up with suitable quality
demineralised water.
The user can decide at his/her own risk to proceed with sterilization.
5.8.3 SUPPLY WATER QUALITY
Only optimal quality demineralised water should be used.
The use of municipal water line water compromises B-One function, causing calcium
build-up, energy dissipation and system malfunction.
The manufacturer declines all responsibility derived from a failure to use quality water.
See the following table for recommended supply water values (ref. "APPENDIX C" norm
UNI EN 13060).
DRY RESIDUE
SILICON OXIDE SiO
IRON
CADMIUM
LEAD
HEAVY METAL
Revision n.1
fig. 20 – icon signalling unsuitable quality of loaded water
RECOMMENDED SUPPLY WATER FEATURES
2
User Manual B-One
≤ 10 mg / l
≤ 1 mg / l
≤ 0.2 mg / l
≤ 0.005 mg / l
≤ 0.05 mg / l
≤ 0.1 mg / l
Edition: 04/2009