FCC Warnings and Notes
• Warning: Changes or modifications to this unit, not expressly approved by Zoetis, could
void the user's authority to operate the equipment.
• Note: This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide
reasonable production against harmful interference when the equipment is operated in a
commercial environment. Operation of this equipment in a residential area is likely to cause
harmful interference in which case the user will be required to correct the interference, at
his own expense.
• Note regarding FCC compliance. Although this instrument has been tested and found to
comply with Part 15, Subpart B of the FCC Rules for a Class A digital device, please note
that this compliance is voluntary, in regard to the cited FCC regulations in effect at the
time of manufacture.
• Note regarding Canadian EMC compliance: This digital apparatus does not emit radio
noise exceeding the limits for class A digital apparatus prescribed in the Radio Interference
Regulations made by the Department of Communications of Canada.
Battery
The VETSCAN UA Analyzer uses a 3.7 V lithium ion rechargeable cell battery to run the system
in the event of AC power loss. When not connected to the AC Charger, the UA will automatically
operate by battery power, the LED will be off and the battery icon will appear to indicate the
charge on the battery. When connected to AC Charger, the Red LED will illuminate while the
battery is charging and turn off after the battery is fully charged. The plug icon on the screen
header will continue to indicate that the UA is operating off AC power.
The battery should fully charge within 5 hours when it is new. A full battery charge will allow
analyzer testing for up to 200 tests. After repeated charging, the battery may lose its ability to
hold a charge.
If less than 100 tests are received on a full charge, it may be an indication that the battery is
getting weak. If this occurs, the analyzer may be run off AC power. Contact Zoetis Technical
Support (See "Zoetis Resource" section) for assistance in replacing the battery.
Warning! Contact Zoetis technical support when attempting to replace the battery. Do not
attempt to change the battery on without Zoetis support.
Perchlorate material – lithium battery contains perchlorate. Special handling may apply, see
www.dtsc.ca.gov/hazardouswaste/perchlorate.