5.
When testing is completed, compare the onscreen results with those on the label on the
Check Strip tube. (Fig. 22) If the results match, the analyzer is fit for analyzing samples.
If the results do not match, run the second Check Strip (provided) to be sure the first strip
is not damaged.
6.
Replace Check Strip into the tube for future use. Be sure that it is not contaminated with
any urine residue from the Strip Tray before placing into the tube. Check Strips may be used
indefinitely, as long as they are not damaged or discolored.