Technical information
Error messages
The measured temperature is above 43.99°C and is
therefore outside the measuring range.
The measured temperature is below 32.00°C and is
therefore outside the measuring range.
The ambient temperature is above 40.00°C and is there-
fore outside the permissible operating conditions.
The ambient temperature is below 10.00°C and is there-
fore outside the permissible operating conditions.
Fever warning. Temperature higher than 37.5°C meas-
ured.
Device memory is full. Connect the ther-
mometer to the cyclotest mySense app to
transfer the measured values to the app.
Battery low. Have a spare battery ready.
Battery flat. Please change the battery.
No display after switching on. Check whether the battery
is positioned correctly and the polarity (+/-) is correct. If
the error persists, change the battery.
EN-18
Bluetooth connection error
Sensor error. If this occurs repeatedly, please contact the
cyclotest service hotline.
Measurement was cancelled prematurely. Repeat the
measurement and wait until the measurement is compete
before removing the thermometer from the measurement
point.
Calibration error. If this occurs repeatedly, please contact
the cyclotest service hotline.
The device is in technical test mode. Switch it off and on
again after a few seconds to return to normal mode.
Technical data
Type:
Maximum thermometer, direct mode
Measuring points/
rectal, oral, vaginal
body reference points:
Temperature unit:
°C
Measuring range:
32.00°C to 43.99°C
Measurement accu-
± 0.05°C between 35.00°C and 38.00°C
racy:
± 0.10°C in other temperature ranges
Memory:
30 measurement values with date and time
can be recalled in memory mode
Operating conditions:
Ambient temperature 10°C to 40°C, maximum
relative humidity 15% to 95%
Technical information
EN-19